Equipment and Calibration History
History for Vaisala GMP222 K1350101
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5003 | 2020-12-01 | Vaira CO2 probe North Profile -0.04 | Removed | Lab none | The battery bank has died and it has been decided to focus on the forced diffusion soil CO2 and Eosense GP profile. The rest of the soil co2 equipment will be removed. |
3123 | 2017-10-05 | Lab none | Installed | Vaira CO2 probe North Profile -0.04 | Installed at Vaira in the North profile at the -4cm location |
3024 | 2017-08-25 | Vaira CO2 probe West Profile -0.08 | Removed | Lab none | Removed from Vaira from West profile -8 cm position for calibration |
2179 | 2016-02-23 | Lab none | Installed | Vaira CO2 probe West Profile -0.08 | Installed in the Vaira West profile at -8cm after lab calibration. |
2153 | 2016-02-04 | Vaira CO2 probe West Profile -0.08 | Removed | Lab none | Removed from Vaira West profile at -8cm for calibration |
2063 | 2014-12-09 | Vaira CO2 probe West Profile -0.08 | Location | Vaira CO2 probe West Profile -0.08 | Vaira West profile at -8cm |
2011 | 2014-04-22 | Vaira CO2 probe West Profile -0.02 | Installed | Lab none | Installed in the Varia West profile at -2cm replacing sn G0230062 |
1447 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.014 + 51.82; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |