Equipment and Calibration History
History for Vaisala GMP222 K1310104
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6027 | 2022-10-05 | Tonzi-SoilCO2 CO2 Probe Open Profile -0.02 | Removed | Lab none | The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -2cm |
3054 | 2017-09-14 | Lab none | Installed | Tonzi CO2 Probe Open Profile -0.02 | Installed at Tonzi in the Open profile at -2cm after lab calibration. This is a new installation with new holes dug. |
3016 | 2017-08-25 | Tonzi CO2 Probe Tree2 Profile -0.04 | Removed | Lab none | Removed from Tonzi Tree 2 profile at -4 cm position for calibration |
2195 | 2016-02-23 | Lab none | Installed | Tonzi CO2 Probe Tree2 Profile -0.04 | Installed in the Tonzi Tree 2 profile at -4cm after lab calibration. |
2159 | 2016-02-04 | Tonzi CO2 Probe Open Profile -0.08 | Removed | Lab none | Removed from Tonzi Open profile at -8cm for calibration |
2072 | 2014-12-09 | Tonzi CO2 Probe Open Profile -0.08 | Location | Tonzi CO2 Probe Open Profile -0.08 | Tonzi Open profile at -8cm |
2015 | 2014-04-08 | Lab none | Installed | Tonzi CO2 Probe Open Profile -0.02 | Installed at Tonzi in the Open profile at -2cm. not clear if it replaced another probe. |
1448 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.008 + 43.09; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |