Field Notes

<--2003-09-11 00:00:00 | 2003-10-02 05:30:00-->
Other sites visited today: Tonzi

Vaira: 2003-09-18 09:00:00 (DOY 261)
Author: Ted Hehn
Others:

Summary: Replace PC, lost communication between data loggers.

When: Fri, 18Sept03, JD-261
Where: Tonzi/Vaira Ranches
Who: Ted Hehn
Weather Clear and hot.

Trip Purpose: Replace PC at Vaira, Repair CO2 profile system and soil temp profile at Tonzi.

Vaira Ranch:

Arrive 09:00, clear and warm

Need to paint fiber glass enclosures. They are beginning to deteriorate and give you glass splinters when you touch them.

Remove PC (Dilbert) and install Hobbs

Communication with 10x & 23x OK

Watch dog timer does not function reliably, begin diagnostics: there is no continuity between the CR23x ground and the flux system ground- for the watch dog to function the lead that goes into the laptop must be grounded to the system ground so if there is no continuity between the 23x ground and the system ground it cannot work.

Place a bonding wire between the ground steak and the data logger battery negative post and watchdog begins to function reliably. However, the intermittent ground that we are experiencing will return at some point and the new ground wire I installed will constitute a ground loop.

I also suspect this intermittent ground is the source of the error we are seeing when the 4 way radiometer parameters switch output order!!!!!!!!!!!! A complete revamping of the Vaira site will be necessary to eliminate this intermittent grounding problem

Blew a 5 amp fuse during power up. Replace 5 amp fuse with 30 amp fuse.

Lost communication with the data loggers and was unable to restore. Believe the registry file is corrupt and will not allow PC208 to access the *.DLL files. Data will not be auto transferred to the laptops so a manual download will need be performed upon next visit.

Field Data

No tank data

No VR Well data found

No TZ Moisture Point data found

No VR Moisture Point data found

No TZ grass heigth found

No VR grass heigth found

No VR Tree DBH found

No water potential data found

No VR Grass Biomass data found

No VR OakLeaves Biomass data found

No VR PineNeedles Biomass data found

0 photos found

Graphs display:
9 sets found

16 VR_met graphs found

Explore the graph:Temperatures in a new window

Explore the graph:System Power Voltage in a new window

Explore the graph:Soil Temperatures Set 1 in a new window

Explore the graph:Soil Temperatures Set 2 in a new window

Explore the graph:Soil Temperature Set 3 in a new window

Explore the graph:Soil Temperature at -4cm in a new window

Explore the graph:Soil Heat Flux in a new window

Explore the graph:Soil Moisture in a new window

Explore the graph:Photosynthetic Active Radiation in a new window

Explore the graph:Short/Long Wave Radiation in a new window

Explore the graph:LED sensor in a new window

Explore the graph:Vaira PPT in a new window

Explore the graph:Peltier SHF Plates in a new window

Explore the graph:Peltier SHF Temperatures in a new window

Explore the graph:LED NDVI in a new window

Explore the graph:NDVI SRS Raw Data in a new window

No data found for VR_flux.
No data found for VR_soilco2.
No data found for VR_pond.
Can't check TZVR_mixed
No data found for VR_fd.
No data found for VR_processed.
No data found for VR_arable.
No data found for VR_soilvue.