Equipment and Calibration
View History
Equipment Entry
|
Documents
Calibration: cal_Vaisala_HMP155_snR0350206_20190118.pdfReport: HMP155 lab and Oxford test - 2019-10-01
Report: SW AirT/RH comparison 2020 - 2020-09-10
Report: WP Golden Set comparison 2020 - 2020-11-19
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5730 | 2022-03-11 | Factory none | Disposed | Disposed none | Campbell Sci inspected the sensor and found that it was shorted and can't be repaired. Discarded at the factory. |
5714 | 2022-03-04 | Lab none | Shipped | Factory none | Shipped back to CSI for inspection and repair. AirT was reading NAN and the AirT chip was very hot to the touch (had slightly melted the white plastic cap) |
5708 | 2022-03-01 | West Pond RH & Temp | Removed | Lab none | Removed from WP and brought back to the lab for further inspection. RH seemed ok but AirT was consistently NAN. It was replaced by HMP155 sn T2820183 the following week. |
5050 | 2021-01-17 | Cal Kit RH & Temp | Moved | West Pond RH & Temp | Moved from Comparison Set to tower to replace sn G4540025 which had high RH readings. |
4975 | 2020-11-19 | Lab none | Installed | Cal Kit RH & Temp | Installed at WP for cross comparison with existing HMP60 which has been reporting low RH |
4948 | 2020-11-12 | Sherman Wetland | Removed | Lab none | Removed from SW and brought back to the lab because the comparison has ended. This reference sensor will be moved to WP next. |
4737 | 2020-09-10 | Lab none | Installed | Sherman Wetland RH & Temp Test_A | Added to the railing near the radiation boom in a ventilated solar shield to compare the new globe thermometers to. |