Equipment and Calibration
View History
Equipment Entry
Instrument: | Precipitation |
Manufacture: | Texas Electronics |
Model No: | TR-525I-R1 |
Serial No: | 24872-999 |
Tag No: | |
Acquisition Date: | |
Project: | Startup |
Location on 2021-10-06: | Tonzi, |
Notes: |
|
|
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
5478 |
2021-10-06 |
Tonzi |
Calibration |
Tonzi none |
Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 54 tips – I turned the screws two turns CCW |
4124 |
2019-10-24 |
Tonzi |
Calibration |
Tonzi |
I connected the bucket to a CR23x dripping 475ml of water through the bucket which should result in 100tips. I got 100 tips. |
3736 |
2018-10-17 |
Tonzi |
Calibration |
Tonzi |
Got 88 of 100 tips - adjusted screws 2.5 turns CW |