Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5295 | 2021-06-02 | Tonzi-Tower Interface LI-7550 | Upgrade | Tonzi-Tower Interface LI-7550 | The firmware was upgraded from v8.5.0 to v8.8.32 for standardization with the Delta Licor systems |
3973 | 2019-04-30 | Tonzi-Floor Interface LI-7500 | Moved | Tonzi-Tower Gas Analyzer CO2/H2O | Moved to the Tower to replace sn AIU-1866 which had some kind of power problem and failed. |
3370 | 2018-03-06 | Tonzi-Floor Interface LI-7500 | Upgrade | Tonzi-Floor Interface LI-7500 | Downgraded from embedded software version 8.7.5 to version 8.5.0 to standardize our systems. We had to downgrade other interfaces to 8.5.0 to solve the data-header mismatch we had with version 8.7.5. |
2968 | 2017-08-17 | Tonzi-Floor | Upgrade | Tonzi-Floor | The firmware was upgraded from v8.5.0 to v8.7.5 |
2764 | 2017-05-11 | Lab none | Installed | Tonzi-Floor | Installed at Tonzi Floor tower. Most of the interface settings seemed to be correct (update rate, data stored, CO2/H2O separation), but we could not figure out how to save the sonic analog settings. The Li-7500 v 8.5 software was cut off on our computer monitor. |