Rec No. | Date | Loc From | Title | Loc To | Notes |
6027 |
2022-10-05 |
Tonzi-SoilCO2 CO2 Probe Open Profile -0.02 |
Removed |
Lab none |
The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -2cm |
3054 |
2017-09-14 |
Lab none |
Installed |
Tonzi CO2 Probe Open Profile -0.02 |
Installed at Tonzi in the Open profile at -2cm after lab calibration. This is a new installation with new holes dug. |
3016 |
2017-08-25 |
Tonzi CO2 Probe Tree2 Profile -0.04 |
Removed |
Lab none |
Removed from Tonzi Tree 2 profile at -4 cm position for calibration |
2195 |
2016-02-23 |
Lab none |
Installed |
Tonzi CO2 Probe Tree2 Profile -0.04 |
Installed in the Tonzi Tree 2 profile at -4cm after lab calibration. |
2159 |
2016-02-04 |
Tonzi CO2 Probe Open Profile -0.08 |
Removed |
Lab none |
Removed from Tonzi Open profile at -8cm for calibration |
2072 |
2014-12-09 |
Tonzi CO2 Probe Open Profile -0.08 |
Location |
Tonzi CO2 Probe Open Profile -0.08 |
Tonzi Open profile at -8cm |
2015 |
2014-04-08 |
Lab none |
Installed |
Tonzi CO2 Probe Open Profile -0.02 |
Installed at Tonzi in the Open profile at -2cm. not clear if it replaced another probe. |
1448 |
2014-04-07 |
Lab none |
Calibration |
Lab none |
Trueppm = Probeppm*1.008 + 43.09; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |