Equipment and Calibration History
History for Campbell Sci. CS451 20010046
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
4597 | 2020-07-09 | Loaned | Disposed | Disposed | Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus. |
4567 | 2020-07-09 | Sherman Barn Water Level | Removed | Loaned none | Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus. |
4526 | 2020-02-06 | Sherman Barn Water Level | Notes | Sherman Barn Water Level | Camilo measured water level: Distance from top of well to water level: 88.1cm (2.89ft) Distance from top of well to ground: 24.7cm (0.81ft) Distance from top of well to sensor: ?? Sensor psi reading: 0.653 psi ~ 46cm |
3578 | 2018-07-11 | Lab none | Installed | Sherman Barn Water Level | Installed at Sherman Barn just to the north of the scaffolding. This is the initial set up of the site. |
3111 | 2017-10-04 | East End-Temp3 Water Level | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
2879 | 2017-06-23 | Lab none | Installed | East End-Temp3 Water Level | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
2433 | 2016-08-24 | Factory | Returned | Lab none | Returned from Campbell on RMA 24692 - cleaned, tested ok and calibrated. |
2446 | 2016-07-28 | Factory | Calibration | Factory | Campbell Sci "Tested unit, it has calcium on the element and needs [to be] cleaned. Cleaned and ran through the environmental calibration chamber. Unit meets specs" |
2358 | 2016-07-13 | Lab none | Shipped | Factory | Sent to Campbell on RMA 24692 to be evaluated. |
2327 | 2016-06-24 | Mayberry Water Level | Removed | Lab none | Removed from the Mayberry tower for testing. Replaced with new probe from DFW system |
859 | 2012-11-29 | Lab none | Installed | Mayberry Water Level | Installed on the Tule tower to replace sn20010026 for repair. |