Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP222
Serial No:H1630066
Tag No:
Acquisition Date:2012-05-23
Project:USDA
Location on 2015-06-09:Lab,
Notes: Came with circuit board GMB220ACB sn H1410159.

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
1887 2015-06-09 Twitchell Alfalfa Removed Lab none Removed from forced diffusion soil cell with upgrade of soil CO2 system
1869 2014-08-13 Lab none Installed Twitchell Alfalfa Installed near the radiometers at Alfalfa in the forced diffusion soil cell
1818 2014-08-07 Sherman Island Removed Lab none Removed with the rest of the soil CO2 system from the Sherman site to be redeployed at the Alfalfa site on Twitchell
1457 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*1.746 + -27.77; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
2087 2014-04-10 Sherman Island Removed Lab none Removed from Sherman profile A at -2cm, replaced with sn K1340005
1353 2014-03-19 Lab none Installed Sherman Island This 0-1% probes was installed at Sherman in Set A at 2cm depth.
1437 2014-01-27 Lab none Calibration Lab none Trueppm = Probeppm*1.753 + -29.94; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty.
958 2013-04-04 Lab none Maintenance Lab none
957 2013-03-28 Mayberry CO2 Probe B Removed Lab none We removed the probe since it was reading 5000mV even we were put the sensor in the air.
907 2013-01-17 Lab none Installed Mayberry CO2 Probe B Installed in a teflon sock in the water in the tule patch
1419 2013-01-08 Lab none Calibration Lab none Trueppm = Probeppm*0.9828 + 15.78; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 421ppm, 3056ppm, and 10000ppm. Ta=22.3C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara.