Equipment and Calibration

View History

Equipment Entry

Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-6400
Serial No:PSC-0785
Tag No:991009170
Acquisition Date:2000-01-31
Project:
Location on 2024-11-13:Lab,
Notes: Portable Photosynthesis System. Leaf Chamber Fluorometer head model 6400-40 serial number LCF-0464

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
7020 2024-11-13 Lab none For Resuse Lab none Marked for reuse
5992 2022-09-13 Borrowed none Returned Lab none LI-6400, all components, and batteries were returned from Rizwan.
5910 2022-07-20 Lab none Loaned Loaned none Loaned to Rizwan Rasheed from Ellen Simms' lab for the summer. Other items loaned: 2 9Ah batteries and a 12V power cord. They had originally borrowed Dave Ackerly's 6400 but we couldn't get it to turn on (display remained off).
5911 2019-06-03 Lab none Loaned Loaned none Loaned to Yvonne Socolar. Other material loaned:
-10 cartridges
-Box of spare parts
-Two 12V Battery
3895 2019-03-25 Lab none Calibration Lab none Camilo Rey-Sanchez and Daphne Szutu

Camilo cleaned the light-source leaf chamber (it wasn't too dusty) and attached it to this console for calibration. We connected the reference and sample inlet to the cal gases with a y-junction. Camilo had checked the zeros using the chemicals last week and found them unsatisfactory (we didn't write them down). We set the zero and the span using cal gas and dewpoint generator.

Zero:
Cal Gas (zero air), flow at 1SCFH:
Before: CO2_S:~-15 CO2_R: ~-15 H2O_S: ~-2 H2O_R: ~-2
After: *All fluctuating around zero.

Span with AMP 408.97ppm cal gas, flow at 1SCFH:
Before: CO2_R: 413 CO2_S: 423
After: *Fluctuating around 409

Span with 16.0C dew point (2019-03-26), flow at 500 CC/min:
Before: H2O_R: 17.50 C H2O_S: 18.52 C
After: H2O_R: 16.00 C H2O_S: 16.00 C
3599 2018-07-16 Lab none Calibration Lab none Elke Eichelmann & Katrina Cone

The 6400 had not been zeroed and spanned for a while. We checked the zero with the chemicals and they were very off, so we set the zero and the span using cal gas and dewpoint generator.

Then I zeroed and spanned:
Zero:
Chemicals: CO2_S: 34.0 CO2_R: 25.4 H2O_S: 13.7 H2O_R: 14.1
Cal Gas (zero air):
Before: CO2_S:41.0 CO2_R: 24.3 H2O_S: 3.5 H2O_R: 2.3
After: *All fluctuating around zero.

Span with 408.97ppm cal gas:
Before: CO2_R: 399.9 CO2_S: 397.1
After: *Fluctuating around 409

Span with 16.0C dew point:
Before: H2O_R: 14.69C H2O_S:13.23C
After: H2O_R: 16.01C H2O_S: 15.99C
**Note: we had to override the maximum slope parameter for the H2O_S to be able to set the slope above 1.100 and to set the correct dewpoint. 'Shift' + 'J' will allow you to override the maximum slope.
2224 2016-04-04 Lab none Calibration Lab none Patty Oikawa

6400 was giving "IRGA's not ready" error in field
Discovered mirrors needed to be cleaned
AGC scores-- Before cleaning: CO2_S:4937 H2O_S:-1500
After cleaning: CO2_S=2652 H2O_S=399.7
*protocol for cleaning mirrors is in the cabinet in downstairs lab

Then I zeroed and spanned:
Zero:
Before: CO2_S:13.8 H2O_S:-0.997
After: CO2_S:0.0 H2O_S:0.003

Span with 421.22ppm cal gas:
Before: CO2_S: 425.4
After: CO2_S: 421.1 (setting 0.998)

Span with 11.08C dew point:
Before: H2O_S:12.18
After: H2O_S: 11.08 (setting 1.039)

1929 2015-08-18 Lab none Calibration Lab none Zeroed and Spanned in lab
CO2 before zero = -24.7uml
after = 0.0uml
H2O before zero = -0.252mml
after = -0.002mml

Ref gas=421.22ppm
CO2 before span = 426.6ppm
after = 421.9ppm

Ref Dew point = 11.08C
H2O before span = 10.49C
after = 11.08C
1734 2015-03-03 Lab none Calibration Lab none Patty calibrated the LI6400, zero and span.
Zero
CO2: -22uml (after calibration: 0.0uml )
H2O: -1.04mml (after calibration: 0.003mml)

Span
Gas = 421.11ppm
CO2: 427.8uml (after calibration: 421.0uml)
Dew point gen=15.95 degrees
H2O: 16.27 (after calibration: 15.95)
1467 2014-06-25 Lab none Calibration Lab none Madeleine and I calibrated the 6400 in the lab
Zero: CO2 1.45 to 0.0
H2O -0.21 to -0.001

Span CO2 with 421.22 CO2 gas
--did not need to span--left at 421.0

H2O span (dew point set at 15.26C)
14.63 to 15.26C
1297 2014-05-12 Lab none Calibration Lab none Zeroed and spanned H2O in the lab
Span: H2O went from dew point 15.56 degrees C to 15.26
Zero: H2O went from 0.1 to 0.0.

Also re-zeroed CO2 for instruction purposes
Zero: CO2 went from 2.1 to 0.0.
1296 2014-05-09 Lab none Calibration Lab none Zeroed and spanned CO2 on IRGA
Zeroes H2O in IRGA
CO2 zeroed from 32 to 0
H2O zeroed 4 to 0
CO2 spanned from 2208.5 to 2238ppm
1093 2013-11-05 Lab none Calibration Lab none Zeroed and spanned H2O and CO2 on IRGA
Neither was far off; CO2 spanned from 419ppm to 421.2ppm
H2O spanned from 14.98C to 15.16C using Dew point generator.
1092 2013-11-05 Lab none Calibration Lab none Zeroed and spanned H2O and CO2 on IRGA
Neither was far off; CO2 spanned from 419ppm to 421.2ppm
H2O spanned from 14.98C to 15.16C using Dew point generator.
1042 2013-08-16 Lab none Notes Lab none Current firmware is OPEN v5.2.3. Upgrade to new firmware v6.2 is possible. New digital board is also available with Ethernet and CF interfaces see Licor part 6400-926 XT Upgrade Kit. Cost at this time is 50.00
757 2012-08-15 Factory Returned Lab none Returned from repair and calibration at Licor on RMA 170405. Licor fixed a intermittent power switch, found a D-shaped vinyl gasket on the sample chamber not installed correctly, replaced leaking home made tube plugs, and found a small leak in a loose bottom cap on the desicant tube. Licor calibrated the system.

software version v5.3.2
758 2012-07-13 Lab none Shipped Factory Shipped to Licor for repair and calibration on RMA 170405. The system has a leak that we have been unable to find.
2466 2007-02-20 Factory Calibration Factory Sent to Licor to check for flow restriction in chemical tubes, test PAR IN stability and calibrate light source. Findings: No CO2 source cover sent. In checking for blockages in the chemical tubes, initially there were none, so I had them rebuilt. In that process we discovered that the small tubes in the bypass valve had been pinched flat. Most likely, they had been left in either the scrub or bypass position for an extended period of time, causing them to be flattened-shut. We replaced them. It is a good idea to out the valve in the center, most relaxed, position for storage. The light source is stable and I could not get it to do anything unusual, even with lots of cable stressing/wiggling.