View history entry 1092 for Licor LI-6400 sn: PSC-0785

Date: 2013-11-05
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Zeroed and spanned H2O and CO2 on IRGA Neither was far off; CO2 spanned from 419ppm to 421.2ppm H2O spanned from 14.98C to 15.16C using Dew point generator.