Equipment and Calibration History

View Equipment |

View history entry 1467 for Licor LI-6400 sn: PSC-0785

Date: 2014-06-25
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Madeleine and I calibrated the 6400 in the lab Zero: CO2 1.45 to 0.0 H2O -0.21 to -0.001 Span CO2 with 421.22 CO2 gas --did not need to span--left at 421.0 H2O span (dew point set at 15.26C) 14.63 to 15.26C
Entered by: Patty Oikawa