Equipment and Calibration History

View Equipment |

View history entry 757 for Licor LI-6400 sn: PSC-0785

Date: 2012-08-15
From: Factory -
To: Lab - none
Title: Returned
Notes: Returned from repair and calibration at Licor on RMA 170405. Licor fixed a intermittent power switch, found a D-shaped vinyl gasket on the sample chamber not installed correctly, replaced leaking home made tube plugs, and found a small leak in a loose bottom cap on the desicant tube. Licor calibrated the system. software version v5.3.2
Entered by: