Rec No. | Date | Loc From | Title | Loc To | Notes |
6897 |
2024-10-14 |
Lab none |
Removed |
Lab none |
Recovered from the bike package. This sensor went to Burning Man in 2023 and has been sitting idle in Robert's office ever since. Stored in equipment cabinet. |
6307 |
2023-07-05 |
Loaned none |
Received |
Lab none |
Returned to the lab from Sophie Ruehr at the conclusion of her measurements along with CR23X sn 2303 |
6229 |
2023-04-05 |
Lab none |
Loaned |
Loaned none |
Loaned to Sophie Ruehr with 23x sn 2303 for field measurements (light extinction curves at the Lake Merrit bonsai garden) |
6208 |
2023-03-07 |
Loaned none |
Returned |
Lab none |
Returned by Sophie Ruehr from lab measurements of grow lights |
6207 |
2023-03-01 |
Lab none |
Loaned |
Loaned none |
Loaned to Sophie Ruehr with 23x sn 2303 for lab measurements of grow lights |
6050 |
2022-10-31 |
Lab none |
Tested |
Lab none |
This sensor was removed because there were spikes in its data on two different mux ports. I pulled apart the splice in the cable and didn't see any problems. From 2022-10-27 to 2022-10-31 I ran an inter-comparison out the lab window with sn 010164. The relationship was tight (r2=0.9999) and the slope was about 1.06. sn 030382 had been inter-compared on the Tonzi tower recently but in the reflected position so the range was not optimal. I think both sensors are functional and should be inter-compared with a good reference in the field to determine appropriate coefficients. |
6037 |
2022-10-20 |
Tonzi PAR Sensor Reflected |
Removed |
Lab none |
Removed from Tonzi due to random spikiness. Replaced by sn 990025. |
6343 |
2021-04-01 |
Tonzi PAR Sensor Reflected |
Location |
Tonzi PAR Sensor Reflected |
Location for sensors currently at Tonzi |
5226 |
2004-04-30 |
Lab none |
Installed |
Tonzi PAR Sensor Reflected Cal |
Installed on the radiation boom for comparison with existing Licor Quantum sensor |