Equipment and Calibration
View History
Equipment Entry
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5362 | 2021-07-06 | Borrowed none | Returned | Lab none | Returned from FLEXLAB at LBL by Hershow Al-Barazi - Cable to be returned later hopefully. |
4477 | 2020-05-22 | Lab none | Loaned | Loaned none | Loaned to Rod Mobini <rmobini@lbl.gov> at LBL FLEXLAB for an experiment on the indoor spread of droplets related to COVID-19 over the summer. The sonic was cleaned and tested okay in the lab. The settings were standardized to: B3 = 9600 baud E1 = Full duplex F1 = 8 data, no parity G1 = No averaging J2 = Two differential inputs J5 = Zero single ended inputs L1 = <cr><lf> M1 = UVW ASCII message NA = Name 'A' O1 = comma separated P6 = 10Hz U1 = m/s Z1 = Not used X1 = align to axis |
1909 | 2015-07-20 | East End Anemometer | Removed | Lab none | Removed from East End to be replaced by a 1590 model sonic with analog outs |
1706 | 2015-01-20 | Lab none | Installed | East End Anemometer | Reinstalled after being removed when the scaffolding fell over. |
1691 | 2015-01-05 | East End Anemometer | Removed | Lab none | Removed from the site due to the tower falling over in the wind on 2014-12-30. Likely undamaged. |
1155 | 2013-11-25 | DWR Corn Anemometer | Moved | East End Anemometer | The DWR Corn / Construction tower was decommissioned and rechristened at its new permanent location at the Twitchell East End Wetland |
736 | 2012-05-21 | Lab none | Installed | DWR Corn | Installed on the DWR Corn scaffolding with the rest of the equipment. |
594 | 2012-05-07 | Twitchell Rice Anemometer | Removed | Lab none | Removed from the tower. Replaced by the new DWR WM (sn ) with analog outs. |
1470 | 2009-04-03 | Lab none | Installed | Twitchell Rice Anemometer | Installed at the Rice. Z= 3.05 m at middle of path & North spar pointing to 318 magnetic.
Need to apply Schotanus correction since model 1352. |
4913 | 2003-05-30 | Lab none | Installed | Tonzi-Floor Anemometer Floor | Installed on Tonzi-Floor to replace sn 000019 that was having problems - not sending data. |
4827 | 2002-04-11 | Vaira Anemometer | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 000028. |
2099 | 2001-01-05 | Lab none | Installed | Vaira Anemometer | Installed on the Vaira Tower to replace sn 000019 that was going off scale at midday. |