Field Notes
<--2025-05-08 10:05:00 | 2025-05-23 09:30:00-->Other sites visited today: Rush Ranch | Gilbert Tract
Phenocam link: Searching...
Hill Slough: 2025-05-22 13:38:00 (DOY 142)
Author: Inbar Amitay
Others: Patty Oikawa, Adan Barrera, Brandon Broach
Summary: Downloaded miniDOT data and changed desiccant, downloaded GHG data, downloaded phenocam, changed 7700 heater setting to summer hours, cleaned sensors, refilled water reservoir.
- Very very windy - DI water reservoir was fully empty - staff gage 4.5m (13:57) - MiniDOT halt record at 13:58, Record at 14:07 Pre clean 1. CO2 -- 423.6 ppm, singal strength 99.3%, H2O 12.2 mmol/mol, Temp 23.09 C, Pressure 101.6 kPa 2. CH4 -- 1.9 umol/mol, singal strength 2.2%, Temp 23.6 C, Pressure 101.2 kPa Post clean 1. CO2 -- 425.5 ppm, H2O 12.1 mmol/mol, Pressure 101.6 kPa, Temp 23.1 C, missing post clean signal strength 2. CH4 -- 1.9 umol/mol, signal strength 80.1% Heater setting - changed start from 16:00 to 20:00 and stop from 10:00 to 08:00 |
0 photos found
4 sets found
21 HS_met graphs found
4 HS_flux graphs found
1 HS_cam graphs found
Can't check HS_Processed
2 HS_Processed graphs found