Field Notes

<--2025-05-08 09:10:00 | 2025-05-23 08:30:00-->
Other sites visited today: Hill Slough | Gilbert Tract
Phenocam link: Searching...

Rush Ranch: 2025-05-22 11:00:00 (DOY 142)
Author: Inbar Amitay
Others: Patty Oikawa, Adan Barrera, Brandon Broach

Summary: Downloaded GHG, phenocam, couldn't connect to met box. Cleaned sensors and topped off reservoir. Measured water well and made sure pressure transducer was below water level.

Pre clean numbers
1. CO2 (7500) -- 423.7 ppm, H2O 11.91 mmol/mol, Temp 23.5 C, Pressure 101.32 kPA, signal strength 94.3%
2. CH4 (7700) -- 2.01 umol/mol, Pressure 100.8 kPa, Temp 23.2 C, signal strength 41%

Post clean numbers
1. CO2 -- 429.6 ppm, signal strength 94.9%
2. CH4 -- 2.006 umol/mol, signal strength 73%

Water well info:
- total length approx 1.54 m until reached soil
- water at approx 40cm above soil
- pressure transducer at sufficient depth, sitting approx 20cm above soil

Left at 13:00 PST

0 photos found

Graphs display:
3 sets found

1 RR_cam graphs found

Explore the graph:GCC in a new window

10 RR_met graphs found

Explore the graph:Air Pressure in a new window

Explore the graph:Temperature in a new window

Explore the graph:Temperature Profile A in a new window

Explore the graph:Temperature Profile B in a new window

Explore the graph:Temperature Profile C in a new window

Explore the graph:Soil Heat Flux in a new window

Explore the graph:Four channel Radiometer in a new window

Explore the graph:Rush Ranch PAR and SW in a new window

Explore the graph:Battery Voltage and RH in a new window

Explore the graph:Soil Moisture in a new window

4 RR_flux graphs found

Explore the graph:CO2 and H2O in a new window

Explore the graph:Wind in a new window

Explore the graph:Methane in a new window

Explore the graph:Data Counts in a new window