26Feb08 16:46 Power on new hi volume pump when unconnected from laser and no particulate matter is observed in exhaust stream. Install new pump into laser system and power up. A momentary cloud of particulate matter is observed in the exhaust stream. Little flow through system. CH4 reading unstable and around 500 ppm. Analyzer failure. Remove analyzer to return to lab. Remove sonic power supply from 120 volt ac and place on 12 volt dc supply and noise in Licor7500 analog traces disappears. |