View history entry 205 for Texas Electronics TR-525I-R1 sn: 24875-999

Date: 2010-09-09
From: Lab - none
To: Lab - none
Title: Calibration
Notes: During the troubleshooting, I ran 16oz (473.3ml) of water through the tipping bucket. According to the manual, this should have resulted in 100 tips. Instead I got 88 tips. I then ran 23.7ml of water that should have resulted in 5 tips and got about 4.5. I then turned the adjustment screws on the bottom two full turns clockwise and got better than five tips. I backed off the screws counter-clockwise a half turn. Now running 473.3ml of water through the tipping bucket results in 99 tips.