Equipment and Calibration
Equipment history at Bouldin Corn| ALL | | Air Quality Sensor || Anemometer || Barometer || Camera || Data Logger || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Infrared Temperature || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || Range Finder || RH & Temp || Soil Moisture Probe || Solar Controller || Water Level || Weather Station |
SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) ORDER BY b.equip, a.eh_date, field(a.eh_title,'Calibration', 'Removed', 'Installed')
Date | Serial No. | Model | Action | Ref | Notes |
---|---|---|---|---|---|
Air Quality Sensor | |||||
2019-02-12 | 68c63acc07a9 | PA-II-SD | Installed | Air Quality Sensor | Installed at BC. It is bolted to the bottom side of the upper boardwalk. |
2019-04-11 | 68c63acc07a9 | PA-II-SD | Removed | Air Quality Sensor | Removed from BC and brought back to the lab for testing. The microSD card was cracked, but the sensor itself was fine. |
2019-04-18 | 68c63acc07a9 | PA-II-SD | Installed | Air Quality Sensor | Installed again at BC after checking the new microSD card and the sensor were ok. |
2019-12-02 | 68c63acc07a9 | PA-II-SD | Removed | Air Quality Sensor | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-17 | 84f3ebd5ffb1 | PA-II-SD | Installed | Air Quality Sensor | Installed at Bouldin Corn after the tower was rebuilt after being knocked down by the wind. The older Purple Air is still in the lab drying out and needs to be put back together. |
2020-02-12 | 84f3ebd5ffb1 | PA-II-SD | Removed | Air Quality Sensor | Brought back to the lab for inspection b/c the wifi network was not coming up |
2020-02-20 | 68c63acc07a9 | PA-II-SD | Installed | Air Quality Sensor | Installed at BC in an additional PVC solar/weather shield and mounted on the southwest corner post of the scaffolding. |
2020-07-21 | 68c63acc07a9 | PA-II-SD | Removed | Air Quality Sensor | Removed from BC and brought back to the lab to test. No data files on card. |
2020-08-21 | 84f3ebd5ffb1 | PA-II-SD | Installed | Air Quality Sensor | Installed at BC on the SW corner of the tower |
2020-09-23 | 84f3ebd5ffb1 | PA-II-SD | Removed | Air Quality Sensor | Removed from BC and brought back to the lab for testing because it does not consistently have files on the card |
2020-09-23 | 68c63acc07a9 | PA-II-SD | Installed | Air Quality Sensor | Installed at BC to replace sn 84f3ebd5ffb1, which was brought back to the lab for testing |
2020-10-22 | 68c63acc07a9 | PA-II-SD | Removed | Air Quality Sensor | Removed from BC and brought back to the lab. The microSD card is hard to seat properly into its slot. It was replaced by sn 84f3ebd5ffb1. |
2020-10-22 | 84f3ebd5ffb1 | PA-II-SD | Installed | Air Quality Sensor | Installed at BC to replace sn 68c63acc07a9, which was brought back to the lab. |
2021-01-29 | 84f3ebd5ffb1 | PA-II-SD | Removed | Air Quality Sensor | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Got rained on and the power connector is corroded |
2021-02-03 | 68c63acc07a9 | PA-II-SD | Installed | Air Quality Sensor | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
2021-11-18 | 68c63acc07a9 | PA-II-SD | Removed | Air Quality Sensor | Removed from BC tower and brought back to the lab because there was some corrosion on the SD card slot |
2021-12-14 | 84f3ebd5ffb1 | PA-II-SD | Installed | Air Quality Sensor | Installed at Bouldin Corn on the southwest scaffolding post. It is powered by mini-USB-to-USB cable that connects in the dfPAR CR10X enclosure. The microSD card slot is tucked up into the inner PVC shield to prevent water from dripping on it. The card slot is easy to pull out and tuck back in when I need to access the card. |
2022-10-18 | 84f3ebd5ffb1 | PA-II-SD | Removed | Air Quality Sensor | Removed from BC tower and brought back to the lab for further inspection. Data over the summer seemed spiky. |
2022-11-02 | 84f3ebd5ffb1 | PA-II-SD | Installed | Air Quality Sensor | Installed at BC on the tower |
2023-11-20 | 84f3ebd5ffb1 | PA-II-SD | Removed | Air Quality Sensor | Removed from BC tower and brought back to the lab. It needs a new data card slot. The current card slot has a bad spring mechanism so cards don't physically mount. |
2023-12-07 | 68c63acc07a9 | PA-II-SD | Installed | Air Quality Sensor | Installed on the BC tower in its usual spot on the southwest corner. This PA has a new card slot that Joe put in. I had brought the sensor home last week so that it could reset its clock on my wifi. However, now campus has an IOT wifi network, so maybe I can try connecting the PA to it next time. |
Anemometer | |||||
2017-04-26 | 074417 | WM 1590 | Installed | Anemometer | Installed at 5.2m agl at the Bouldin Corn site with the initial set up of the tower. |
2019-12-02 | 074417 | WM 1590 | Removed | Anemometer | One traducer missing. Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | 153805 | WM 1590 | Installed | Anemometer | Installed at Bouldin Corn after re-assembling tower. This unit replaces sn 074417 that lost a transducer when the tower fell. |
2021-01-29 | 153805 | WM 1590 | Removed | Anemometer | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Partially submerged |
2021-02-03 | 153803 | WM 1590 | Installed | Anemometer | Installed on the BC tower to replace sensors damaged when the tower fell last week |
2024-02-14 | 022365410 | SONIC-ANEMO-SPE10 | Installed | none | Installed at Bouldin Corn with a new, udpated Mark 3, sn D006617. |
Barometer | |||||
2017-04-26 | L2340502 | PTB110 | Installed | Barometer | Installed at the Bouldin Corn site with the initial set up of the tower. |
2019-12-02 | L2340502 | PTB110 | Removed | Barometer | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | L2340502 | PTB110 | Installed | Barometer | Re-installed at Bouldin Corn after re-assembling tower. |
2021-12-01 | L2340502 | PTB110 | Removed | Barometer | Removed from BC because it was giving bad data of 80 kPa and brought back to the lab for testing |
Camera | |||||
2017-07-12 | 0030F4D21491 | SD130BN | Installed | Camera | This camera was put up on the Bouldin Corn Tower. It was attached to the southwest corner of the scaffolding just below the top railing and connected the to NL115 on the CR1000 to record the images on a CFM card. The CR1000 IP was set to 192.168.2.100 and the FTP user=Tower and pw=biomet. |
2019-12-02 | 0030F4D21491 | SD130BN | Removed | Camera | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | 0030F4D21491 | SD130BN | Installed | Camera | Re-installed at Bouldin Corn after re-assembling tower. |
2020-05-19 | RPI0W-001 | RPI0W | Installed | none | Installed at BC on the ground in a normal-density corn patch, looking upwards. |
2020-05-19 | RPI0W-002 | RPI0W | Installed | none | Installed at BC on the ground in a normal-density corn patch, looking upwards. |
2020-05-19 | RPI0W-004 | RPI0W | Installed | none | Installed at BC on the ground in a normal-density corn patch, looking upwards. This camera has a regular lens, and we will compare the photos with 2 picams with fisheye lenses. |
2020-07-30 | RPI0W-001 | RPI0W | Removed | Removed from BC and brought back to the lab because the corn canopy is about done developing for the season | |
2020-07-30 | RPI0W-002 | RPI0W | Removed | Removed from BC and brought back to the lab because the corn canopy is about done developing for the season | |
2020-07-30 | RPI0W-004 | RPI0W | Removed | Removed from BC and brought back to the lab because the corn canopy is about done developing for the season | |
2021-01-29 | 0030F4D21491 | SD130BN | Removed | Camera | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
2021-02-03 | 0030F4D21491 | SD130BN | Installed | Camera | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
2021-05-27 | RPI0W-001 | RPI0W | Installed | none | Installed under the corn canopy at BC. This camera is in line with a row. |
2021-05-27 | RPI0W-002 | RPI0W | Installed | none | Installed under the corn canopy at BC. This camera is between rows. |
2021-08-17 | RPI0W-001 | RPI0W | Removed | Removed from the BC field and brought back to the lab since the canopy has finished developing for this year | |
2021-08-17 | RPI0W-002 | RPI0W | Removed | Removed from the BC field and brought back to the lab since the canopy has finished developing for this year | |
2022-05-27 | RPI0W-001 | RPI0W | Installed | camera B | Installed in line with a corn row. |
2022-05-27 | RPI0W-002 | RPI0W | Installed | Camera C | Installed between corn rows |
2022-06-08 | RPI0W-001 | RPI0W | Removed | camera B | This camera was dead. There was a puddle of water in the bottom of the enclosure and white corrosion all over the board and mini-USB connector. |
2022-08-31 | RPI0W-002 | RPI0W | Removed | Camera C | Removed this zcam from BC and brought back to the lab since the corn canopy has finished development and we won't get any more interesting photos |
2023-05-31 | RPI0W-002 | RPI0W | Installed | none | Installed in the corn field at BC for the 2023 growing season |
2023-05-31 | RPI0W-007 | RPI0W | Installed | none | Installed in the corn field at BC for the 2023 growing season |
2023-08-24 | RPI0W-007 | RPI0W | Removed | camera B | Removed from the z-cam location because the corn is finished growing |
2023-08-24 | RPI0W-002 | RPI0W | Removed | Camera C | Removed from the z-cam location because the corn is finished growing |
2024-05-22 | RPI0W-007 | RPI0W | Installed | none | The clock was bad after being off overnight, so I changed the RTC battery (CR1225) and reset the clock. Installed in the Bouldin corn field on the ground to capture gap fraction through the growing season. Tractor work should be done for now and plants are 5-10cm tall. |
2024-05-22 | RPI0W-002 | RPI0W | Installed | none | Installed in the Bouldin corn field on the ground to capture gap fraction through the growing season. Tractor work should be done for now and plants are 5-10cm tall. |
2024-08-30 | RPI0W-007 | RPI0W | Removed | Camera C | Zcam removed from field because sorghum canopy has finished development |
2024-08-30 | RPI0W-002 | RPI0W | Removed | camera B | Zcam removed from field because sorghum canopy has finished development |
Data Logger | |||||
2017-04-26 | 66251 | CR1000 | Installed | Data Logger Met | Installed on the Bouldin Corn Tower to collect the met data with the initial set up of the site. |
2017-07-14 | 31308 | CR10X | Installed | Data Logger Diffuse PAR | Installed to do the high frequency measurements for diffuse PAR and report the CR1000 met data logger. |
2017-07-26 | 31308 | CR10X | Location | Data Logger Diffuse PAR | Check the serial number of the data logger at the site. |
2019-12-02 | 31308 | CR10X | Removed | Data Logger Diffuse PAR | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 66251 | CR1000 | Removed | Data Logger Met | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | 66251 | CR1000 | Installed | Data Logger Met | Re-installed at Bouldin Corn after re-assembling tower. |
2019-12-13 | 31308 | CR10X | Installed | Data Logger Diffuse PAR | Re-installed at Bouldin Corn after re-assembling tower. |
2021-06-09 | 23865 | CR1000 | Moved | none | Moved from BC to HS as part of rad cal kit |
2021-06-16 | 1132 | CR1000X | Installed | Data Logger Surface Renewal | Installed as part of Kevin Tu's surface renewal experiment |
2024-01-04 | 1132 | CR1000X | Removed | Data Logger Surface Renewal | Removed from BC at the conclusion of Corteva's IRT experiment. |
Forced Diffusion | |||||
2019-11-14 | FD20180067 | eosFD | Installed | Forced Diffusion C | Installed in the corn field in the unharvested section |
2019-11-14 | FD20180048 | eosFD | Installed | Forced Diffusion B | Installed in the corn field near Tyler's southern soil gas ports |
2019-11-14 | FD20180047 | eosFD | Installed | Forced Diffusion A | Installed in the corn field near Tyler's northern soil gas ports |
2019-12-20 | FD20180047 | eosFD | Removed | Forced Diffusion A | Removed from the field by Tyler due to being partially submerged by winter flooding. |
2019-12-20 | FD20180048 | eosFD | Removed | Forced Diffusion B | Removed from the field by Tyler due to being partially submerged by winter flooding. |
2019-12-20 | FD20180067 | eosFD | Removed | Forced Diffusion C | Removed from the field by Tyler due to being partially submerged by winter flooding. |
2020-01-07 | FD20180047 | eosFD | Installed | Forced Diffusion A | Reinstalled at Bouldin Corn after being removed when the field was flooded. Installed near previous location on 12 inch collars. |
2020-01-07 | FD20180048 | eosFD | Installed | Forced Diffusion B | Reinstalled at Bouldin Corn after being removed when the field was flooded. Installed near previous location on 12 inch collars. |
2020-01-07 | FD20180067 | eosFD | Installed | Forced Diffusion C | Reinstalled at Bouldin Corn after being removed when the field was flooded. Installed near previous location on 12 inch collars. |
2020-04-17 | FD20180067 | eosFD | Moved | Forced Diffusion C | Removed from the corn field and installed under the tower because there will be tractor work on the field soon. |
2020-04-17 | FD20180047 | eosFD | Removed | Forced Diffusion A | Removed from BC and brought back to the lab because there will be tractor work on the field soon. |
2020-04-17 | FD20180048 | eosFD | Removed | Forced Diffusion B | Removed from BC and brought back to the lab because there will be tractor work on the field soon. |
2020-05-19 | FD20180067 | eosFD | Moved | Forced Diffusion C | Moved back into the field from being under the scaffolding at the side of the field for tilling and planting. This chamber is still on a tall collar and is in normal density corn. |
2020-05-29 | FD20180047 | eosFD | Installed | Forced Diffusion A | Reinstalled in normal density corn just in front of tower |
2020-05-29 | FD20180048 | eosFD | Installed | Forced Diffusion B | Reinstalled in normal density corn just in front of tower |
2020-06-09 | FD20180047 | eosFD | Removed | Forced Diffusion A | Removed from the field because it was not responding. Cable tested okay. Maybe go too hot. |
2020-07-21 | FD20180047 | eosFD | Installed | Forced Diffusion A | Installed in the normal density corn in front of the tower. |
2020-09-01 | FD20180048 | eosFD | Moved | Forced Diffusion B | Moved tower and soil sensors back from field in preparation for corn harvest. Chambers were reinstalled under the tower. |
2020-09-01 | FD20180067 | eosFD | Moved | Forced Diffusion C | Moved tower and soil sensors back from field in preparation for corn harvest. Chambers were reinstalled under the tower. |
2020-09-01 | FD20180047 | eosFD | Removed | Forced Diffusion A | Removed for testing in the lab because it was not responding in the field. Probably its high temperature non-start problem. |
2020-09-10 | FD20180047 | eosFD | Installed | Forced Diffusion A | Reinstalled after having been returned to the lab when it would not start due to being hot. (Note by Daphne 2021-08-11: I'm pretty sure this chamber was installed under the tower since we were waiting for our field to be harvested at this time). |
2020-09-23 | FD20180048 | eosFD | Moved | Forced Diffusion B | Moved from BC to BA to replace sn FD20160001, which was brought back to the lab for testing |
2020-09-23 | FD20180047 | eosFD | Removed | Forced Diffusion A | Removed and brought back to the lab for testing because it would not connect in the field |
2021-07-20 | FD20180067 | eosFD | Maintenance | Forced Diffusion C | Performance check done. Referenced reset to 400ppm. Average Flux: -0.4, 0.09 STDEV Average ATM: 551.1ppm, 39.5C Average Soil: 534.8ppm, 39.5C Large difference between internal chambers. |
2022-01-02 | FD20180067 | eosFD | Removed | Forced Diffusion C | Removed from BC underneath the tower and brought back to the lab for testing. It was 1/2 submerged under water starting 2021-12-25, and Tyler put it on the lower boardwalk on 2021-12-29. |
2022-02-17 | FD20180067 | eosFD | Installed | Forced Diffusion C | Installed at BC under the tower |
2022-06-08 | FD20180067 | eosFD | Removed | Forced Diffusion C | Removed from field and brought back to the lab because even though the pump sounds like it's working, we haven't been able to connect to it for the past month |
2022-07-14 | FD20180067 | eosFD | Installed | Forced Diffusion C | Installed in the corn at BC. It's in the same location from where it was removed, one row west of the zcam. I connected to the FD chamber to confirm that it was running. |
2023-01-03 | FD20180067 | eosFD | Removed | Forced Diffusion C | Removed from BC and brought the chamber and cable back to the lab for testing. We found it floating in the water underneath the tower. The cable connector was underwater. |
2023-03-08 | FD20180026 | eosFD | Installed | Forced Diffusion C | Installed under BC tower to replace sn FD20180067 which got flooded |
2023-05-09 | FD20180026 | eosFD | Moved | Forced Diffusion C | Moved from BC to BA to replace sn FD20160001 which broke. We should have soil CO2 flux data from Whendee's chamber measurements at BC. |
Gas Analyzer | |||||
2017-04-26 | 75H-2182 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at 5.2m agl at the Bouldin Corn site with the initial set up of the tower. |
2017-08-16 | 75H-2182 | LI-7500A | Calibration | Lab | Last calibration: 2017-04-20 AGC=87.5 SDM=0 DAC1=0 DAC2=0 Ambient P=100.4kPa, Tair=23.9C, Tdew=12.7C Zero Air at 1075psi, AMP Air 408.97 at 2000 Tdewspan=16.00 CO2 zero 0.8838 -> .8832 Offset: -2.73ppm CO2 span 1.0164 -> 1.0123 Offset: 2.13ppm H2O zero 0.8454 -> .8436 Offset: -11.4ppm H2O span 1.0139 -> 1.0038 Offset: .2ppm Notes: |
2017-04-26 | TG1-0421 | LI-7700 | Installed | Gas Analyzer CH4 | Installed at 5.2m agl the Bouldin Corn site with the initial set up of the tower. |
2017-07-12 | TG1-0421 | LI-7700 | Removed | Gas Analyzer CH4 | Brought back to the lab to replace its power board. Power board provided by Li-cor. |
2017-07-14 | TG1-0421 | LI-7700 | Installed | Gas Analyzer CH4 | Installed at Bouldin Corn tower |
2017-07-26 | 75H-2182 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Brought down for calibration and replaced by 75H-2668. |
2017-08-16 | 75H-2182 | LI-7500A | Calibration | Lab | Last calibration: 2017-04-20 AGC=87.5 SDM=0 DAC1=0 DAC2=0 Ambient P=100.4kPa, Tair=23.9C, Tdew=12.7C Zero Air at 1075psi, AMP Air 408.97 at 2000 Tdewspan=16.00 CO2 zero 0.8838 -> .8832 Offset: -2.73ppm CO2 span 1.0164 -> 1.0123 Offset: 2.13ppm H2O zero 0.8454 -> .8436 Offset: -11.4ppm H2O span 1.0139 -> 1.0038 Offset: .2ppm Notes: |
2017-07-26 | 75H-2668 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Replaced 75H-2182. |
2017-11-28 | 75H-2668 | LI-7500A | Calibration | Lab | Last calibration: 2017-07-25 AGC=99.9 SDM=0 DAC1=0 DAC2=0 Ambient P=100.6kPa, Tair=21.7C, Tdew=8.9C Zero Air at 300psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.9395 -> .9392 Offset: -1.56ppm CO2 span 1.0159 -> 1.0147 Offset: 0.64ppm H2O zero 0.9047 -> 0.9066 Offset: 11.0ppm H2O span 1.0107 -> 1.0125 Offset: -0.04ppm Notes: Calibrated on 7550 software version 8.7.5 |
2017-10-04 | TG1-0421 | LI-7700 | Removed | Gas Analyzer CH4 | Removed from the Bouldin Corn tower for calibration and testing in the lab for drop outs when washing problem. Replaced by sn TG1-0211. |
2017-10-04 | TG1-0211 | LI-7700 | Installed | Gas Analyzer CH4 | Installed on the Bouldin Corn tower to replace sn TG1-0421 for calibration and testing in the lab for drop outs when washing problem. |
2017-10-19 | 75H-2668 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 1594 |
2017-11-28 | 75H-2668 | LI-7500A | Calibration | Lab | Last calibration: 2017-07-25 AGC=99.9 SDM=0 DAC1=0 DAC2=0 Ambient P=100.6kPa, Tair=21.7C, Tdew=8.9C Zero Air at 300psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.9395 -> .9392 Offset: -1.56ppm CO2 span 1.0159 -> 1.0147 Offset: 0.64ppm H2O zero 0.9047 -> 0.9066 Offset: 11.0ppm H2O span 1.0107 -> 1.0125 Offset: -0.04ppm Notes: Calibrated on 7550 software version 8.7.5 |
2017-10-19 | 1594 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed on Bouldin Corn to replace sn 75H-2668 for calibration. |
2017-11-13 | 1594 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn because it was calibrated on LI-7550 with embedded software v. 8.7.5 and the LI-7550 at the tower is embedded software v 8.5.0, so we were worried this software mismatch would produce wonky data. Replaced by sn 75H-2176. |
2017-11-13 | 75H-2176 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed on Bouldin Corn replacing sn 1594 for calibration. |
2018-02-05 | 75H-2176 | LI-7500A | Calibration | Factory | Sensor calibrated at the Li-cor factory. CO2 Calibration Values A= 1.58409E2 B= -1.53312E4 C= 6.48285E7 D= -2.17327E10 E= 2.80839E12 XS= 0.0004 Z= -1.4E-3 SD1= -0.0085 SD2= -0.0892 SD3= 2.0680 H2O Calibration Values A= 5.83891E3 B= 4.03548E6 C= -9.20453E6 XS= -0.0013 Z= 2.1E-2 SD1= 0.0484 SD2= 1.3610 SD3= 2.3810 Signal Strength B= 2.158 C= 4.122 |
2018-01-04 | 75H-2176 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from tower because of PLL error (did not hear sensor fan spinning) |
2018-02-05 | 75H-2176 | LI-7500A | Calibration | Factory | Sensor calibrated at the Li-cor factory. CO2 Calibration Values A= 1.58409E2 B= -1.53312E4 C= 6.48285E7 D= -2.17327E10 E= 2.80839E12 XS= 0.0004 Z= -1.4E-3 SD1= -0.0085 SD2= -0.0892 SD3= 2.0680 H2O Calibration Values A= 5.83891E3 B= 4.03548E6 C= -9.20453E6 XS= -0.0013 Z= 2.1E-2 SD1= 0.0484 SD2= 1.3610 SD3= 2.3810 Signal Strength B= 2.158 C= 4.122 |
2018-01-16 | 0035 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at BC to replace sn 75H-2176 (had PLL error) |
2018-06-25 | 0035 | LI-7500A | Calibration | Lab | Last calibration: 2018-01-12 AGC=99.2 SDM=0 DAC1=- DAC2=- Ambient P=100.5kPa, Tair=25.8C, Tdew=12.7C Zero Air at 350psi, AMP 408.97 at 1350 Tdewspan=15.99 CO2 zero 0.9378 -> 0.9374 Offset: -1.8ppm CO2 span 1.0288 -> 1.0296 Offset: 5.83ppm H2O zero 0.8978 -> 0.8979 Offset: -0.1ppm H2O span 1.0296 -> 1.0261 Offset: 0.08ppm Notes: Calibrated by Joe |
2018-04-05 | 0035 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from BC and brought back to the lab for calibration. It was replaced by sn 75H-2668. |
2018-06-25 | 0035 | LI-7500A | Calibration | Lab | Last calibration: 2018-01-12 AGC=99.2 SDM=0 DAC1=- DAC2=- Ambient P=100.5kPa, Tair=25.8C, Tdew=12.7C Zero Air at 350psi, AMP 408.97 at 1350 Tdewspan=15.99 CO2 zero 0.9378 -> 0.9374 Offset: -1.8ppm CO2 span 1.0288 -> 1.0296 Offset: 5.83ppm H2O zero 0.8978 -> 0.8979 Offset: -0.1ppm H2O span 1.0296 -> 1.0261 Offset: 0.08ppm Notes: Calibrated by Joe |
2018-04-05 | TG1-0211 | LI-7700 | Removed | Gas Analyzer CH4 | Removed from BC and brought back to the lab for calibration. it was replaced by TG1-0155. |
2018-04-05 | 75H-2668 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at BC to replace sn 0035, which was brought back to the lab for calibration. |
2018-09-26 | 75H-2668 | LI-7500A | Calibration | Lab | Last calibration: 2018-04-04 AGC=100.0 SDM=0 DAC1=0 DAC2=0 Ambient P=100.4kPa, Tair=26.5C, Tdew=13.8C Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=16.00 CO2 zero 0.9393 -> 0.9392 Offset: -.27ppm CO2 span 1.0147 -> 1.0153 Offset: -.27ppm H2O zero 0.9038 -> .9056 Offset: 10.6ppm H2O span 1.0146 -> 1.0166 Offset: .05ppm Notes: "Check PLL" error for the first 30 seconds, then status changed to "OK." |
2018-04-05 | TG1-0155 | LI-7700 | Installed | Gas Analyzer CH4 | Installed at BC to replace sn TG1-0211, which was brought back to the lab for calibration. |
2018-07-12 | 75H-2668 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from BC and brought back to the lab for calibration. It was replaced by sn 75H-2182. |
2018-09-26 | 75H-2668 | LI-7500A | Calibration | Lab | Last calibration: 2018-04-04 AGC=100.0 SDM=0 DAC1=0 DAC2=0 Ambient P=100.4kPa, Tair=26.5C, Tdew=13.8C Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=16.00 CO2 zero 0.9393 -> 0.9392 Offset: -.27ppm CO2 span 1.0147 -> 1.0153 Offset: -.27ppm H2O zero 0.9038 -> .9056 Offset: 10.6ppm H2O span 1.0146 -> 1.0166 Offset: .05ppm Notes: "Check PLL" error for the first 30 seconds, then status changed to "OK." |
2018-07-12 | 75H-2182 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at BC to replace sn 75H-2668, which was brought back to the lab for calibration. |
2018-10-24 | 75H-2182 | LI-7500A | Calibration | Lab | Last calibration: 2018-07-10 AGC=87.3 SDM=0 DAC1=0 DAC2=0 Ambient P=100.55kPa, Tair=26.87C, Tdew=12.61C Zero Air at 1500psi, AMP Air 408.97 at 1100 Tdewspan=16.00 CO2 zero 0.8841 -> .8844 Offset: 1.11ppm CO2 span 1.0168 -> 1.0188 Offset: -1.04ppm H2O zero 0.845 -> .8449 Offset: -.6ppm H2O span 1.0202 -> 1.0143 Offset: .13ppm Notes: |
2018-10-18 | 75H-2182 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 75H-2668. |
2018-10-24 | 75H-2182 | LI-7500A | Calibration | Lab | Last calibration: 2018-07-10 AGC=87.3 SDM=0 DAC1=0 DAC2=0 Ambient P=100.55kPa, Tair=26.87C, Tdew=12.61C Zero Air at 1500psi, AMP Air 408.97 at 1100 Tdewspan=16.00 CO2 zero 0.8841 -> .8844 Offset: 1.11ppm CO2 span 1.0168 -> 1.0188 Offset: -1.04ppm H2O zero 0.845 -> .8449 Offset: -.6ppm H2O span 1.0202 -> 1.0143 Offset: .13ppm Notes: |
2018-10-18 | 75H-2668 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed on Bouldin Corn to replace sn 75H-2182 for calibration |
2019-02-11 | 75H-2668 | LI-7500A | Calibration | Lab | Last calibration: 2018-10-08 AGC=99 SDM=0 DAC1=0 DAC2=0 Ambient P=101.1kPa, Tair=22.5C, Tdew=3.7C Zero Air at 1150psi, AMP Air 408.97 at 800 Tdewspan=16.00 CO2 zero 0.9394 -> .939 Offset: -1.51ppm CO2 span 1.0164 -> .018 Offset: -.81ppm H2O zero 0.9057 -> .9073 Offset: 8.8ppm H2O span 1.0157 -> 1.0117 Offset: 0.08ppm Notes: Dewpoint generator dewpoint kept falling b/c it was low on water. I added more DI water and it seemed fine. |
2018-10-25 | TG1-0155 | LI-7700 | Calibration | Gas Analyzer CH4 | Calibration Location: Bouldin Corn Last calibration: 2018-04-04 RSSI=91%, Optical RH=16% Ambient P=101.5kPa, Tair=27.2C Zero Air at 1925psi, AMP Air 3.121 at 1975 CH4 Zero Offset: .015ppm CH4 Span Offset: -.013ppm Notes: Gently breezy day |
2019-01-31 | 75H-2668 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from BC and brought back to the lab for calibration. It was replaced by sn 1594. |
2019-02-11 | 75H-2668 | LI-7500A | Calibration | Lab | Last calibration: 2018-10-08 AGC=99 SDM=0 DAC1=0 DAC2=0 Ambient P=101.1kPa, Tair=22.5C, Tdew=3.7C Zero Air at 1150psi, AMP Air 408.97 at 800 Tdewspan=16.00 CO2 zero 0.9394 -> .939 Offset: -1.51ppm CO2 span 1.0164 -> .018 Offset: -.81ppm H2O zero 0.9057 -> .9073 Offset: 8.8ppm H2O span 1.0157 -> 1.0117 Offset: 0.08ppm Notes: Dewpoint generator dewpoint kept falling b/c it was low on water. I added more DI water and it seemed fine. |
2019-01-31 | 1594 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at BC to replace sn 75H-2668, which was brought back to the lab for calibration. |
2019-04-11 | TG1-0155 | LI-7700 | Calibration | Gas Analyzer CH4 | Calibration Location: Bouldin Corn Last calibration: 2018-10-25 RSSI=87%, Optical RH=22% Ambient P=101.5kPa, Tair=19.1C Zero Air M4002 at 1600psi, AMP C03923 3.121 at 1600 CH4 Zero Offset: 0.024ppm CH4 Span Offset: -0.007ppm Notes: Windy |
2019-04-25 | TG1-0155 | LI-7700 | Upgrade | Gas Analyzer CH4 | The LI-7700 firmware was upgraded from v1.0.21 to v1.0.27 mostly minor bug fixes see change log https://www.licor.com/env/support/LI-7700/topics/change-log.html |
2019-04-25 | 1594 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 0073. |
2019-04-25 | 0073 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed on Bouldin Corn to replace sn 1594 for calibration |
2019-08-14 | 0073 | LI-7500A | Calibration | Lab | Last calibration: 2019-04-23 AGC=98.7 SDM= DAC1= DAC2= Ambient P=100.2kPa, Tair=29.6C, Tdew=13.2C Zero Air at 1000psi, AMP CA4396 408.97 at 850 Tdewspan=16.00 CO2 zero 0.9563 -> .9576 Offset: 4.5ppm CO2 span 1.0094 -> 1.0095 Offset: -.13ppm H2O zero 0.7336 -> .7318 Offset: -10.5ppm H2O span 1.0153 -> 1.0247 Offset: -.5ppm Notes: When the SS was 98.7, I reset it to 100. The previous calibration on 2019-04-23, the H2O zero was offset by 12.1 mmol/m3, and this time the offset was -10.5 mmol/m3. |
2019-08-07 | 0073 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from BC and brought back to the lab for calibration. It was replaced by sn 75H-2669. |
2019-08-14 | 0073 | LI-7500A | Calibration | Lab | Last calibration: 2019-04-23 AGC=98.7 SDM= DAC1= DAC2= Ambient P=100.2kPa, Tair=29.6C, Tdew=13.2C Zero Air at 1000psi, AMP CA4396 408.97 at 850 Tdewspan=16.00 CO2 zero 0.9563 -> .9576 Offset: 4.5ppm CO2 span 1.0094 -> 1.0095 Offset: -.13ppm H2O zero 0.7336 -> .7318 Offset: -10.5ppm H2O span 1.0153 -> 1.0247 Offset: -.5ppm Notes: When the SS was 98.7, I reset it to 100. The previous calibration on 2019-04-23, the H2O zero was offset by 12.1 mmol/m3, and this time the offset was -10.5 mmol/m3. |
2019-08-07 | 75H-2669 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at BC to replace sn 0073, which was brought back to the lab for calibration. |
2020-02-04 | 75H-2669 | LI-7500A | Calibration | Lab | Last calibration: 2019-08-01 AGC=102.0 SDM=0 DAC1=- DAC2=- Ambient P=101.3kPa, Tair=25.6C, Tdew=-5.2C Zero Air at 850psi, AMP CA04396 408.97 at 550 Tdewspan=16.0 CO2 zero 0.9178 -> 0.9184 Offset: 2.3ppm CO2 span 1.0125 -> 1.0160 Offset: -1.87ppm H2O zero 0.9019 -> 0.9006 Offset: -8.4ppm H2O span 1.015 -> 1.0101 Offset: 0.1ppm Notes: |
2019-10-22 | TG1-0155 | LI-7700 | Calibration | Gas Analyzer CH4 | Calibration Location: Bouldin Corn Last calibration: 2019-04-11 RSSI=89.5%, Optical RH=26% Ambient P=102.1kPa, Tair=17.2C Zero Air at 1350psi, AMP CA03923 3.121 at 1400 CH4 Zero Offset: 0.024ppm CH4 Span Offset: -0.005ppm Notes: Light breeze, swapped optical desiccant after calibration using cal gas to dry head space. New RH 7%. |
2019-11-14 | 75H-2669 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn and brought back to lab for calibration. Replaced by sn 0042. |
2020-02-04 | 75H-2669 | LI-7500A | Calibration | Lab | Last calibration: 2019-08-01 AGC=102.0 SDM=0 DAC1=- DAC2=- Ambient P=101.3kPa, Tair=25.6C, Tdew=-5.2C Zero Air at 850psi, AMP CA04396 408.97 at 550 Tdewspan=16.0 CO2 zero 0.9178 -> 0.9184 Offset: 2.3ppm CO2 span 1.0125 -> 1.0160 Offset: -1.87ppm H2O zero 0.9019 -> 0.9006 Offset: -8.4ppm H2O span 1.015 -> 1.0101 Offset: 0.1ppm Notes: |
2019-11-14 | 0042 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 75H-2669 for calibration. |
2019-12-04 | 0042 | LI-7500A | Calibration | Lab | Last calibration: 2019-11-12 AGC=100.5 SDM= DAC1= DAC2= Ambient P=100.1kPa, Tair=20.3C, Tdew=10.11C Zero Air at 850psi, AMP CA4396 408.97 at 550 Tdewspan=20.00 CO2 zero 0.9481 -> .9490 Offset: 3.19ppm CO2 span 1.0091 -> 1.0182 Offset: -4.69ppm H2O zero 0.8449 -> .8441 Offset: -4.7ppm H2O span 1.0139 -> 1.0341 Offset: -.45ppm Notes: This sensor was on the BC tower when it fell down because of a wind/rain storm on 2019-11-26. First calibration since it was brought back to the lab. The CO2 span is larger than I expected. More testing needed before putting it back up in the field. |
2019-12-02 | TG1-0155 | LI-7700 | Removed | Gas Analyzer CH4 | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 0042 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-04 | 0042 | LI-7500A | Calibration | Lab | Last calibration: 2019-11-12 AGC=100.5 SDM= DAC1= DAC2= Ambient P=100.1kPa, Tair=20.3C, Tdew=10.11C Zero Air at 850psi, AMP CA4396 408.97 at 550 Tdewspan=20.00 CO2 zero 0.9481 -> .9490 Offset: 3.19ppm CO2 span 1.0091 -> 1.0182 Offset: -4.69ppm H2O zero 0.8449 -> .8441 Offset: -4.7ppm H2O span 1.0139 -> 1.0341 Offset: -.45ppm Notes: This sensor was on the BC tower when it fell down because of a wind/rain storm on 2019-11-26. First calibration since it was brought back to the lab. The CO2 span is larger than I expected. More testing needed before putting it back up in the field. |
2019-12-13 | TG1-0155 | LI-7700 | Installed | Gas Analyzer CH4 | Re-installed at Bouldin Corn after re-assembling tower. |
2019-12-13 | 0042 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Re-installed at Bouldin Corn after re-assembling tower. |
2020-05-13 | 0042 | LI-7500A | Calibration | Lab | Last calibration: 2019-12-04 AGC=99.9 SDM= DAC1= DAC2= Ambient P=100.9kPa, Tair=19.9C, Tdew=9.18C Zero Air at 750psi, AMP CA4396 408.97 at 550 Tdewspan=16.00 CO2 zero 0.949 -> .9481 Offset: -3.44ppm CO2 span 1.0182 -> 1.0141 Offset: 2.03ppm H2O zero 0.8441 -> .8441 Offset: 0.1ppm H2O span 1.0341 -> 1.0271 Offset: 0.15ppm Notes: |
2020-02-20 | 0042 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 75H-2668. |
2020-05-13 | 0042 | LI-7500A | Calibration | Lab | Last calibration: 2019-12-04 AGC=99.9 SDM= DAC1= DAC2= Ambient P=100.9kPa, Tair=19.9C, Tdew=9.18C Zero Air at 750psi, AMP CA4396 408.97 at 550 Tdewspan=16.00 CO2 zero 0.949 -> .9481 Offset: -3.44ppm CO2 span 1.0182 -> 1.0141 Offset: 2.03ppm H2O zero 0.8441 -> .8441 Offset: 0.1ppm H2O span 1.0341 -> 1.0271 Offset: 0.15ppm Notes: |
2020-02-20 | 75H-2668 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 0042 for calibration. |
2020-05-20 | 75H-2668 | LI-7500A | Calibration | Lab | Last calibration: 2020-02-11 AGC=100.3 SDM=0 DAC1=- DAC2=- Ambient P=100.9kPa, Tair=20.4C, Tdew=9.7C Zero Air at 800psi, AMP CA04396 408.97 at 600 Tdewspan=16.0 CO2 zero 0.9393 -> 0.9390 Offset: -1.3ppm CO2 span 1.0171 -> 1.0198 Offset: -1.37ppm H2O zero 0.9072 -> 0.9082 Offset: 5.7ppm H2O span 1.0147 -> 1.0119 Offset: 0.05ppm Notes: |
2020-05-19 | TG1-0155 | LI-7700 | Calibration | Calibration Location: Bouldin Corn Last calibration: 2019-12-04 RSSI=88%, Optical RH=9% Ambient P=101.6kPa, Tair=17.5C Zero Air at 900psi, AMP CA3923 3.121 at 1000 CH4 Zero Offset: 0.022ppm CH4 Span Offset: -0.001ppm Notes: Seems stable (2nd calibration after sensor fell off BC tower). Breezy day. | |
2020-05-19 | 75H-2668 | LI-7500A | Removed | Removed from Bouldin Corn and brought back to the lab for calibration. It was replaced by sn 75H-2176. | |
2020-05-20 | 75H-2668 | LI-7500A | Calibration | Lab | Last calibration: 2020-02-11 AGC=100.3 SDM=0 DAC1=- DAC2=- Ambient P=100.9kPa, Tair=20.4C, Tdew=9.7C Zero Air at 800psi, AMP CA04396 408.97 at 600 Tdewspan=16.0 CO2 zero 0.9393 -> 0.9390 Offset: -1.3ppm CO2 span 1.0171 -> 1.0198 Offset: -1.37ppm H2O zero 0.9072 -> 0.9082 Offset: 5.7ppm H2O span 1.0147 -> 1.0119 Offset: 0.05ppm Notes: |
2020-05-19 | 75H-2176 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 2668 for calibration. |
2020-08-25 | 75H-2176 | LI-7500A | Calibration | Lab | Last calibration: 2020-05-13 AGC=100.7 SDM=0 DAC1=- DAC2=- Ambient P=100.2kPa, Tair=23.8C, Tdew=14.7C Zero Air at 700psi, AMP CA04396 408.97 at 450 Tdewspan=20.0 CO2 zero 0.8672 -> 0.8671 Offset: -0.1ppm CO2 span 1.0081 -> 1.0055 Offset: 1.33ppm H2O zero 0.8521 -> 0.8526 Offset: 2.9ppm H2O span 0.9159 -> 0.9246 Offset: -0.2ppm Notes: |
2020-08-21 | 75H-2176 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 0073. |
2020-08-25 | 75H-2176 | LI-7500A | Calibration | Lab | Last calibration: 2020-05-13 AGC=100.7 SDM=0 DAC1=- DAC2=- Ambient P=100.2kPa, Tair=23.8C, Tdew=14.7C Zero Air at 700psi, AMP CA04396 408.97 at 450 Tdewspan=20.0 CO2 zero 0.8672 -> 0.8671 Offset: -0.1ppm CO2 span 1.0081 -> 1.0055 Offset: 1.33ppm H2O zero 0.8521 -> 0.8526 Offset: 2.9ppm H2O span 0.9159 -> 0.9246 Offset: -0.2ppm Notes: |
2020-08-21 | 0073 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 75H-2176 for calibration. |
2020-12-08 | 0073 | LI-7500A | Calibration | Lab | Last calibration: 2020-08-20 AGC=99.8 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=24.4C, Tdew=2.3C Zero Air at 550psi, AMP CA4396 408.97 at 300 Tdewspan=20.00 CO2 zero 0.9575 -> .9564 Offset: -3.46ppm CO2 span 1.0092 -> 1.0079 Offset: .42ppm H2O zero 0.7312 -> .7323 Offset: 6.7ppm H2O span 1.0118 -> 1.0037 Offset: .18ppm Notes: |
2020-11-20 | TG1-0155 | LI-7700 | Calibration | Gas Analyzer CH4 | Calibration Location: Bouldin Corn Last calibration: 2020-05-19 RSSI=88%, Optical RH=20% Ambient P=102.5kPa, Tair=9.4C Zero Air at 575psi, AMP CA3923 3.121 at 550 CH4 Zero Offset: 0ppm CH4 Span Offset: 0.009ppm Notes: Foggy day with a litle breeze. The RSSI after cleaning was 50, but jumped up to 88 after putting hood on and running dry air through. |
2020-12-03 | 0073 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 75H-2176. |
2020-12-08 | 0073 | LI-7500A | Calibration | Lab | Last calibration: 2020-08-20 AGC=99.8 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=24.4C, Tdew=2.3C Zero Air at 550psi, AMP CA4396 408.97 at 300 Tdewspan=20.00 CO2 zero 0.9575 -> .9564 Offset: -3.46ppm CO2 span 1.0092 -> 1.0079 Offset: .42ppm H2O zero 0.7312 -> .7323 Offset: 6.7ppm H2O span 1.0118 -> 1.0037 Offset: .18ppm Notes: |
2020-12-03 | 75H-2176 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 0073 for calibration. |
2021-03-01 | 75H-2176 | LI-7500A | Calibration | Lab | Last calibration: 2020-12-02 AGC=102.5 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=22.9C, Tdew=4.6C Zero Air at 450psi, AMP CA04396 408.97 at 250 Tdewspan=20.0 CO2 zero 0.8673 -> 0.8671 Offset: -0.25ppm CO2 span 1.009 -> 1.000 Offset: 2.53ppm H2O zero 0.8531 -> 0.8524 Offset: -1.3ppm H2O span 0.9238 -> 0.9092 Offset: 0.37ppm Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm. |
2021-01-29 | 75H-2176 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Partially submerged |
2021-03-01 | 75H-2176 | LI-7500A | Calibration | Lab | Last calibration: 2020-12-02 AGC=102.5 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=22.9C, Tdew=4.6C Zero Air at 450psi, AMP CA04396 408.97 at 250 Tdewspan=20.0 CO2 zero 0.8673 -> 0.8671 Offset: -0.25ppm CO2 span 1.009 -> 1.000 Offset: 2.53ppm H2O zero 0.8531 -> 0.8524 Offset: -1.3ppm H2O span 0.9238 -> 0.9092 Offset: 0.37ppm Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm. |
2021-01-29 | TG1-0155 | LI-7700 | Removed | Gas Analyzer CH4 | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
2021-02-03 | TG1-0224 | LI-7700 | Installed | Gas Analyzer CH4 | Installed on the BC tower to replace sensors damaged when the tower fell last week |
2021-02-03 | 1594 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed on the BC tower to replace sensors damaged when the tower fell last week |
2021-03-04 | 1594 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn and brought back to the lab for calibration. Replaced by sn 75H-2514. |
2021-03-04 | 75H-2514 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 1594 for calibration. |
2021-08-31 | 75H-2514 | LI-7500A | Calibration | Lab | Last calibration: 2021-03-01 AGC=101.2 SDM=0 DAC1=- DAC2=- Ambient P=99.7kPa, Tair=26.2C, Tdew=13.1C Zero Air at 1800psi, AMP CA04119 524.77 at 1700 Tdewspan=20.0 CO2 zero 0.9048 -> 0.9046 Offset: -0.45ppm CO2 span 0.9984 -> 1.0082 Offset: -6.77ppm H2O zero 0.8389 -> 0.8399 Offset: 7.1ppm H2O span 1.0061 -> 1.0133 Offset: -0.16ppm Notes: |
2021-05-27 | TG1-0224 | LI-7700 | Calibration | Calibration Location: Bouldin Corn Last calibration: 2020-11-30 RSSI=73%, Optical RH=15% Ambient P=101.6kPa, Tair=26.8C AMP FB03864 at 900psi, AMP CA04119 4.223 at 1900 CH4 Zero Offset: 0.006ppm CH4 Span Offset: -0.003ppm Notes: Zero gas was an AMP cal gas in a baby tank that we borrowed for the day. | |
2021-06-09 | 75H-2514 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 75H-2150. |
2021-08-31 | 75H-2514 | LI-7500A | Calibration | Lab | Last calibration: 2021-03-01 AGC=101.2 SDM=0 DAC1=- DAC2=- Ambient P=99.7kPa, Tair=26.2C, Tdew=13.1C Zero Air at 1800psi, AMP CA04119 524.77 at 1700 Tdewspan=20.0 CO2 zero 0.9048 -> 0.9046 Offset: -0.45ppm CO2 span 0.9984 -> 1.0082 Offset: -6.77ppm H2O zero 0.8389 -> 0.8399 Offset: 7.1ppm H2O span 1.0061 -> 1.0133 Offset: -0.16ppm Notes: |
2021-06-09 | 75H-2150 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 75H-2514 for calibration. |
2021-09-29 | 75H-2150 | LI-7500A | Calibration | Lab | Last calibration: 2021-06-07 AGC=100 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=26.3C, Tdew=9.07C Zero Air at 1425psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.9024 -> .9018 Offset: -2.57ppm CO2 span 1.0107 -> 1.0034 Offset: 5.1ppm H2O zero 0.8687 -> .8689 Offset: 1.1ppm H2O span 1.0111 -> 1.0112 Offset: -.01ppm Notes: |
2021-06-24 | TG1-0357 | LI-7700 | Calibration | Gas Analyzer CH4 | Calibration Location: BC Last calibration: 2021-01-20 RSSI=80.2%, Optical RH=23% Ambient P=101.5kPa, Tair=34.5C Zero Air at 1800psi, AMP CA04119 4.2238 at 1800 CH4 Zero Offset: -0.015ppm CH4 Span Offset: -0.0338ppm Notes: This 7700 was just returned from Nick and Sabina who were using it on Staten. They calibrated it on 2021-01-20 but I don't have notes on that. Also I uploaded the TG1-0224 config file before the calibration and again after calibration but I don't think that has any affect on the calibration or measurements. |
2021-06-24 | TG1-0224 | LI-7700 | Removed | Gas Analyzer CH4 | Removed from BC and loaned to Steve, Sabina and Nick for use on Staten Island. This unit belongs to DWR and the one they had belongs to UCB. |
2021-06-24 | TG1-0357 | LI-7700 | Installed | Gas Analyzer CH4 | Returned from Steve, Sabina and Nick who were using it on Staten Island. This unit belongs to UCB and was swapped for sn TG1-0224 which belongs to DWR. |
2021-09-23 | 75H-2150 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 0041. |
2021-09-29 | 75H-2150 | LI-7500A | Calibration | Lab | Last calibration: 2021-06-07 AGC=100 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=26.3C, Tdew=9.07C Zero Air at 1425psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.9024 -> .9018 Offset: -2.57ppm CO2 span 1.0107 -> 1.0034 Offset: 5.1ppm H2O zero 0.8687 -> .8689 Offset: 1.1ppm H2O span 1.0111 -> 1.0112 Offset: -.01ppm Notes: |
2021-09-23 | 0041 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 75H-2150 for calibration. |
2021-12-16 | 0041 | LI-7500A | Calibration | Lab | Last calibration: 2021-09-21 AGC=100.8 SDM= DAC1= DAC2= Ambient P=100.9kPa, Tair=24.0C, Tdew=6.2C Zero Air at 1000psi, AMP CA04119 524.77 at 1650 Tdewspan=20.00 CO2 zero 0.9667 -> .9960 Offset: -2.19ppm CO2 span 1.0145 -> 1.0172 Offset: -1.79ppm H2O zero 0.8224 -> .8227 Offset: 1.5ppm H2O span 1.0228 -> 1.0335 Offset: -.23ppm Notes: |
2021-12-01 | TG1-0357 | LI-7700 | Calibration | Calibration Location: Bouldin Corn Last calibration: 2021-06-24 RSSI=73%, Optical RH=24% Ambient P=101.7kPa, Tair=20.9C Zero Air at 1100psi, AMP CA04396 1.9317 at 1700 CH4 Zero Offset: 0.021ppm CH4 Span Offset: -.0017ppm Notes: Relatively still day, lots of geese in the air. After calibration, flushed headspace with zero air until optics RH=10%. Swapped out the desiccant bottle for one with fresh drierite. Zero air tank pressure uncertain--wonky regulator. | |
2021-12-14 | 0041 | LI-7500A | Removed | Removed from Bouldin Corn for calibration and brought back to the lab. Replaced by sn 0042. | |
2021-12-16 | 0041 | LI-7500A | Calibration | Lab | Last calibration: 2021-09-21 AGC=100.8 SDM= DAC1= DAC2= Ambient P=100.9kPa, Tair=24.0C, Tdew=6.2C Zero Air at 1000psi, AMP CA04119 524.77 at 1650 Tdewspan=20.00 CO2 zero 0.9667 -> .9960 Offset: -2.19ppm CO2 span 1.0145 -> 1.0172 Offset: -1.79ppm H2O zero 0.8224 -> .8227 Offset: 1.5ppm H2O span 1.0228 -> 1.0335 Offset: -.23ppm Notes: |
2021-12-14 | 0042 | LI-7500A | Installed | none | Installed at Bouldin Corn to replace sn 0041 for calibration. |
2022-04-19 | 0042 | LI-7500A | Calibration | Lab | Last calibration: 2021-12-09 AGC=100.4 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=25.8C, Tdew=7.5C Zero Air at 650psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9481 -> .9483 Offset: .85ppm CO2 span 1.0091 -> 1.0087 Offset: .14ppm H2O zero 0.8452 -> .8444 Offset: -4.9ppm H2O span 1.0213 -> 1.0191 Offset: .05ppm Notes: |
2022-04-13 | 0042 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration and brought back to the lab. Replaced by sn 0065. |
2022-04-19 | 0042 | LI-7500A | Calibration | Lab | Last calibration: 2021-12-09 AGC=100.4 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=25.8C, Tdew=7.5C Zero Air at 650psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.9481 -> .9483 Offset: .85ppm CO2 span 1.0091 -> 1.0087 Offset: .14ppm H2O zero 0.8452 -> .8444 Offset: -4.9ppm H2O span 1.0213 -> 1.0191 Offset: .05ppm Notes: |
2022-04-13 | 0065 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 0042 for calibration. |
2022-09-12 | 0065 | LI-7500A | Calibration | Lab | Last calibration: 2022-04-08 AGC=101.0 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=22.8C, Tdew=15.2C Zero Air at 1900psi, AMP CA04119 524.77 at 1400 Tdewspan=20.0 CO2 zero 0.8908 -> 0.8918 Offset: 4.5ppm CO2 span 1.0198 -> 1.0227 Offset: -1.97ppm H2O zero 0.7308 -> 0.7316 Offset: 6.7ppm H2O span 1.016 -> 1.0167 Offset: -0.01ppm Notes: |
2022-06-08 | TG1-0357 | LI-7700 | Calibration | Calibration Location: Bouldin Corn Last calibration: 2021-12-01 RSSI=77%, Optical RH=16% Ambient P=101kPa, Tair=28C Zero Air at 600psi, AMP CA04119 4.2238 at 1600 CH4 Zero Offset: -.074ppm CH4 Span Offset: .0078ppm Notes: Extra windy day of course. Did not change drierite or flush headspace with zero air. | |
2022-08-31 | 0065 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 75H-2180. |
2022-09-12 | 0065 | LI-7500A | Calibration | Lab | Last calibration: 2022-04-08 AGC=101.0 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=22.8C, Tdew=15.2C Zero Air at 1900psi, AMP CA04119 524.77 at 1400 Tdewspan=20.0 CO2 zero 0.8908 -> 0.8918 Offset: 4.5ppm CO2 span 1.0198 -> 1.0227 Offset: -1.97ppm H2O zero 0.7308 -> 0.7316 Offset: 6.7ppm H2O span 1.016 -> 1.0167 Offset: -0.01ppm Notes: |
2022-08-31 | 75H-2180 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 0065 for calibration. |
2022-12-05 | 75H-2180 | LI-7500A | Calibration | Lab | Last calibration: 2022-08-30 AGC=101.0 SDM=0 DAC1=- DAC2=- Ambient P=100.6kPa, Tair=21.3C, Tdew=9.2C Zero Air at 1700psi, AMP CA04119 524.77 at 1300 Tdewspan=16.0 CO2 zero 0.9148 -> 0.9145 Offset: -1.53ppm CO2 span 1.0318 -> 1.0433 Offset: -7.27ppm H2O zero 0.8916 -> 0.8933 Offset: 10.7ppm H2O span 1.0122 -> 1.0381 Offset: -0.53ppm Notes: Large offset in CO2 span but also large offset when removed from tower and similar offset in sn 2176 CO2 span calibrated at the same time. Used our 7550 box 0674 instead on the AMP loaner used previously. I think this IRGA is okay. |
2022-11-30 | TG1-0357 | LI-7700 | Calibration | Gas Analyzer CH4 | Calibration Location: Bouldin Corn Last calibration: 2022-06-08 RSSI=80%, Optical RH=23% Ambient P=101.3kPa, Tair=12.6C Zero Air at 1650psi, AMP CA04396 1.9317 at 1100 CH4 Zero Offset: .032ppm CH4 Span Offset: .0083ppm Notes: Breezy. After calibration we flushed headspace with zero air until optics RH = 8%. We swapped in fresh drierite. Half of the thermocouple tubing had broken off so we put on a new piece of tubing. |
2022-11-30 | 75H-2180 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 1594. |
2022-12-05 | 75H-2180 | LI-7500A | Calibration | Lab | Last calibration: 2022-08-30 AGC=101.0 SDM=0 DAC1=- DAC2=- Ambient P=100.6kPa, Tair=21.3C, Tdew=9.2C Zero Air at 1700psi, AMP CA04119 524.77 at 1300 Tdewspan=16.0 CO2 zero 0.9148 -> 0.9145 Offset: -1.53ppm CO2 span 1.0318 -> 1.0433 Offset: -7.27ppm H2O zero 0.8916 -> 0.8933 Offset: 10.7ppm H2O span 1.0122 -> 1.0381 Offset: -0.53ppm Notes: Large offset in CO2 span but also large offset when removed from tower and similar offset in sn 2176 CO2 span calibrated at the same time. Used our 7550 box 0674 instead on the AMP loaner used previously. I think this IRGA is okay. |
2022-11-30 | 1594 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 75H-2180 for calibration. |
2023-01-06 | 1594 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn because no CO2/H2O data since mid December. Replaced by sn 0041. |
2023-01-06 | 0041 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 1594 because sn 1594 was not giving any data. |
2023-04-24 | 0041 | LI-7500A | Calibration | Lab | Last calibration: 2023-01-05 AGC=100 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=21.7C, Tdew=6.85C Zero Air at 1300psi, AMP CA04119 524.77 at 1150 Tdewspan=16.00 CO2 zero 0.9961 -> .9968 Offset: 2.54ppm CO2 span 1.0234 -> 1.0202 Offset: 2.12ppm H2O zero 0.8232 -> .8222 Offset: -5.8ppm H2O span 1.0382 -> 1.0302 Offset: .17ppm Notes: Span point entered as "524.67" for sensor to read ~524.77 after spanning |
2023-04-11 | 0041 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 75H-2669. |
2023-04-24 | 0041 | LI-7500A | Calibration | Lab | Last calibration: 2023-01-05 AGC=100 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=21.7C, Tdew=6.85C Zero Air at 1300psi, AMP CA04119 524.77 at 1150 Tdewspan=16.00 CO2 zero 0.9961 -> .9968 Offset: 2.54ppm CO2 span 1.0234 -> 1.0202 Offset: 2.12ppm H2O zero 0.8232 -> .8222 Offset: -5.8ppm H2O span 1.0382 -> 1.0302 Offset: .17ppm Notes: Span point entered as "524.67" for sensor to read ~524.77 after spanning |
2023-04-11 | 75H-2669 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 0041 for calibration. |
2023-08-04 | 75H-2669 | LI-7500A | Calibration | Lab | Last calibration: 2023-04-10 AGC=100.6 SDM= DAC1= DAC2= AIU sn AIU-1569, Presure A0=58.004, A1=15.479 Ambient P=100.6kPa, Tair=22.1C, Tdew=13.9C Zero Air at 900psi, AMP CA04119 524.77 at 875 Tdewspan=16.00 CO2 zero 0.9207 -> .9195 Offset: -4.61ppm CO2 span 1.0176 -> 1.0167 Offset: .58ppm H2O zero 0.9001 -> .9009 Offset: 5.3ppm H2O span 1.0105 -> 1.0163 Offset: -.12ppm Notes: Vertical orientation |
2023-06-14 | TG1-0357 | LI-7700 | Calibration | Calibration Location: Bouldin Corn Last calibration: 2022-11-30 RSSI=75%, Optical RH=23% Ambient P=101.1kPa, Tair=23.0C Zero Air at 1250psi, AMP CA04119 4.2238 at 1125 CH4 Zero Offset: -.012ppm CH4 Span Offset: .0238ppm Notes: Breezy. Flushed headspace to 12% optics RH after calibration and swapped in a new desiccant bottle. | |
2023-07-24 | 75H-2669 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 0418. |
2023-08-04 | 75H-2669 | LI-7500A | Calibration | Lab | Last calibration: 2023-04-10 AGC=100.6 SDM= DAC1= DAC2= AIU sn AIU-1569, Presure A0=58.004, A1=15.479 Ambient P=100.6kPa, Tair=22.1C, Tdew=13.9C Zero Air at 900psi, AMP CA04119 524.77 at 875 Tdewspan=16.00 CO2 zero 0.9207 -> .9195 Offset: -4.61ppm CO2 span 1.0176 -> 1.0167 Offset: .58ppm H2O zero 0.9001 -> .9009 Offset: 5.3ppm H2O span 1.0105 -> 1.0163 Offset: -.12ppm Notes: Vertical orientation |
2023-07-24 | 0418 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 75H-2669 for calibration. |
2023-10-23 | 0418 | LI-7500A | Calibration | Lab | Last calibration: 2023-07-19 AGC=97.3 SDM= DAC1= DAC2= AIU sn AIU-1569, Presure A0=58.004, A1=15.479 Ambient P=100.1kPa, Tair=23.1C, Tdew=29.0C Zero Air at 700psi, AMP CA04119 524.77 at 775 Tdewspan=20.00 CO2 zero 0.907 -> .9072 Offset: .71ppm CO2 span 1.0069 -> 1.0066 Offset: .14ppm H2O zero 0.933 -> .9331 Offset: .6ppm H2O span 1.0049 -> 1.0052 Offset: -.01ppm Notes: Vertical orientation. Changed SS from 97.3 to 100 before calibration. |
2023-10-11 | 0418 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 0042. |
2023-10-23 | 0418 | LI-7500A | Calibration | Lab | Last calibration: 2023-07-19 AGC=97.3 SDM= DAC1= DAC2= AIU sn AIU-1569, Presure A0=58.004, A1=15.479 Ambient P=100.1kPa, Tair=23.1C, Tdew=29.0C Zero Air at 700psi, AMP CA04119 524.77 at 775 Tdewspan=20.00 CO2 zero 0.907 -> .9072 Offset: .71ppm CO2 span 1.0069 -> 1.0066 Offset: .14ppm H2O zero 0.933 -> .9331 Offset: .6ppm H2O span 1.0049 -> 1.0052 Offset: -.01ppm Notes: Vertical orientation. Changed SS from 97.3 to 100 before calibration. |
2023-10-11 | 0042 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 0418 for calibration. |
2024-03-23 | 0042 | LI-7500A | Calibration | Lab | Last calibration: 2023-10-05 AGC=100.8 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.08kPa, Tair=24.9C, Tdew=9.91C Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00 CO2 zero 0.9473 -> .9483 Offset: 3.82ppm CO2 span 1.0033 -> 1.0098 Offset: -4.49ppm H2O zero 0.8474 -> .8451 Offset: -13.6ppm H2O span 1.0075 -> 1.02 Offset: -.28ppm Notes: H2O span was oscillating within 1ppm (but average value was stable), but this went away after adding more DI water to the dewpoint generator reservoir. CO2 span seems larger than usual. Painted some chips in the paint with nail polish |
2024-01-04 | 0042 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 0065. |
2024-03-23 | 0042 | LI-7500A | Calibration | Lab | Last calibration: 2023-10-05 AGC=100.8 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.08kPa, Tair=24.9C, Tdew=9.91C Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00 CO2 zero 0.9473 -> .9483 Offset: 3.82ppm CO2 span 1.0033 -> 1.0098 Offset: -4.49ppm H2O zero 0.8474 -> .8451 Offset: -13.6ppm H2O span 1.0075 -> 1.02 Offset: -.28ppm Notes: H2O span was oscillating within 1ppm (but average value was stable), but this went away after adding more DI water to the dewpoint generator reservoir. CO2 span seems larger than usual. Painted some chips in the paint with nail polish |
2024-01-04 | 0065 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 0042 for calibration. |
2024-04-15 | 0065 | LI-7500A | Calibration | Lab | Last calibration: 2024-01-03 AGC=95.57 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.49 Ambient P=100.86kPa, Tair=21.18C, Tdew=C Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00 CO2 zero 0.8918 -> .8916 Offset: -.84ppm CO2 span 1.0255 -> 1.024 Offset: 1.01ppm H2O zero 0.7329 -> .7329 Offset: -.1ppm H2O span 1.0238 -> 1.021 Offset: .06ppm Notes: Used 524.67 as CO2 span point in software for post-span CO2 value to be ~524.77. |
2024-01-11 | TG1-0357 | LI-7700 | Calibration | Calibration Location: Bouldin Corn Last calibration: 2023-06-14 RSSI=71%, Optical RH=9% Ambient P=102.0kPa, Tair=13.3C Zero Air at 1700psi, AMP CA03923 3.3865 at 1700 CH4 Zero Offset: .035ppm CH4 Span Offset: -.0215ppm Notes: Still day for once | |
2024-03-25 | 0065 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 75H-2180. |
2024-04-15 | 0065 | LI-7500A | Calibration | Lab | Last calibration: 2024-01-03 AGC=95.57 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.49 Ambient P=100.86kPa, Tair=21.18C, Tdew=C Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00 CO2 zero 0.8918 -> .8916 Offset: -.84ppm CO2 span 1.0255 -> 1.024 Offset: 1.01ppm H2O zero 0.7329 -> .7329 Offset: -.1ppm H2O span 1.0238 -> 1.021 Offset: .06ppm Notes: Used 524.67 as CO2 span point in software for post-span CO2 value to be ~524.77. |
2024-03-25 | 75H-2180 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 0065 for calibration. |
2024-09-26 | 75H-2180 | LI-7500A | Calibration | Lab | Last calibration: 2024-03-23 AGC=102.5 SDM=0 DAC1=- DAC2=- AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.0kPa, Tair=24.8C, Tdew=14.3C Zero Air at 1350psi, AMP CA03923 488.75 at 1450 Tdewspan=20.0 CO2 zero 0.9137 -> 0.9139 Offset: 1.1ppm CO2 span 1.0356 -> 1.0373 Offset: -1.05ppm H2O zero 0.8951 -> 0.8954 Offset: 1.3ppm H2O span 1.0393 -> 1.0468 Offset: -0.15ppm Notes: Set the CO2 span using 488.55 instead of 488.75 as this resulted in a 488.75 measurement |
2024-07-22 | 75H-2180 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 0035. |
2024-09-26 | 75H-2180 | LI-7500A | Calibration | Lab | Last calibration: 2024-03-23 AGC=102.5 SDM=0 DAC1=- DAC2=- AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.0kPa, Tair=24.8C, Tdew=14.3C Zero Air at 1350psi, AMP CA03923 488.75 at 1450 Tdewspan=20.0 CO2 zero 0.9137 -> 0.9139 Offset: 1.1ppm CO2 span 1.0356 -> 1.0373 Offset: -1.05ppm H2O zero 0.8951 -> 0.8954 Offset: 1.3ppm H2O span 1.0393 -> 1.0468 Offset: -0.15ppm Notes: Set the CO2 span using 488.55 instead of 488.75 as this resulted in a 488.75 measurement |
2024-07-22 | 0035 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 75H-2180 for calibration. |
2024-08-07 | TG1-0357 | LI-7700 | Calibration | Calibration Location: Bouldin Corn Last calibration: 2024-01-11 RSSI=75%, Optical RH=17% Ambient P=101.2kPa, Tair=31.0C Zero Air at 1700psi, AMP CA03923 3.3865 at 1900 CH4 Zero Offset: -.043ppm CH4 Span Offset: 0.0135ppm Notes: Still day | |
2024-10-23 | 0035 | LI-7500A | Removed | Gas Analyzer CO2/H2O | Removed from Bouldin Corn for calibration. Replaced by sn 0041. |
2024-10-23 | 0041 | LI-7500A | Installed | Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 0035 for calibration. |
Heat Flux Plate | |||||
2017-04-26 | 10867 | HFP01 | Installed | Heat Flux Plate B | Installed at the Bouldin Corn site with the initial set up of the tower. Temporarily at the south edge of the scaffolding until tractor work is finished. |
2017-04-26 | 11326 | HFP01 | Installed | Heat Flux Plate A | Installed at the Bouldin Corn site with the initial set up of the tower. Temporarily at the south edge of the scaffolding until tractor work is finished. |
2017-04-26 | 10861 | HFP01 | Installed | Heat Flux Plate C | Installed at the Bouldin Corn site with the initial set up of the tower. Temporarily at the south edge of the scaffolding until tractor work is finished. |
2019-12-02 | 10861 | HFP01 | Removed | Heat Flux Plate C | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 10867 | HFP01 | Removed | Heat Flux Plate B | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 11326 | HFP01 | Removed | Heat Flux Plate A | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | 11326 | HFP01 | Installed | Heat Flux Plate A | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
2019-12-13 | 10867 | HFP01 | Installed | Heat Flux Plate B | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
2019-12-13 | 10861 | HFP01 | Installed | Heat Flux Plate C | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
2020-05-19 | 11326 | HFP01 | Moved | Heat Flux Plate A | Moved back into the field from being under the scaffolding at the side of the field for tilling and planting. |
2020-05-19 | 10867 | HFP01 | Moved | Heat Flux Plate B | Moved back into the field from being under the scaffolding at the side of the field for tilling and planting. |
2020-05-19 | 10861 | HFP01 | Moved | Heat Flux Plate C | Moved back into the field from being under the scaffolding at the side of the field for tilling and planting. |
2020-09-01 | 11326 | HFP01 | Moved | Heat Flux Plate A | Moved tower and soil sensors back from field in preparation for corn harvest. Soil sensors were reinstalled under the tower at -2cm depth. |
2020-09-01 | 10867 | HFP01 | Moved | Heat Flux Plate B | Moved tower and soil sensors back from field in preparation for corn harvest. Soil sensors were reinstalled under the tower at -2cm depth. |
2020-09-01 | 10861 | HFP01 | Moved | Heat Flux Plate C | Moved tower and soil sensors back from field in preparation for corn harvest. Soil sensors were reinstalled under the tower at -2cm depth. |
2023-04-26 | 10861 | HFP01 | Moved | Heat Flux Plate C | Moved from field to tower in preparation for impending plowing. Not installed in soil yet. |
2023-04-26 | 10867 | HFP01 | Moved | Heat Flux Plate B | Moved from field to tower in preparation for impending plowing. Not installed in soil yet. |
2023-04-26 | 11326 | HFP01 | Moved | Heat Flux Plate A | Moved from field to tower in preparation for impending plowing. Not installed in soil yet. |
2023-05-03 | 10861 | HFP01 | Installed | Heat Flux Plate C | Installed in soil under tower at 2cm depth. |
2023-05-03 | 10867 | HFP01 | Installed | Heat Flux Plate B | Installed in soil under tower at 2cm depth. |
2023-05-03 | 11326 | HFP01 | Installed | Heat Flux Plate A | Installed in soil under tower at 2cm depth. |
Infrared Temperature | |||||
2021-06-16 | 1073 | SI-131 | Installed | Infrared Temperature North | Installed as part of Kevin Tu's surface renewal experiment |
2021-06-16 | 1076 | SI-131 | Installed | Infrared Temperature South | Installed as part of Kevin Tu's surface renewal experiment |
2024-01-04 | 1073 | SI-131 | Removed | Infrared Temperature North | Removed from BC at the conclusion of Corteva's IRT experiment. |
2024-01-04 | 1076 | SI-131 | Removed | Infrared Temperature South | Removed from BC at the conclusion of Corteva's IRT experiment. |
Interface | |||||
2017-04-26 | AIU-1570 | LI-7550 | Installed | Interface LI-7550 | Installed at the Bouldin Corn site with the initial set up of the tower. |
2017-04-26 | 13056 | NL115 | Installed | none | Installed at BC with initial set up of site |
2017-08-23 | AIU-1570 | LI-7550 | Upgrade | Interface LI-7550 | Upgraded the 7550 firmware from v8.0.0 to v8.7.5. I confirmed the 7550 time, time zone, sonic analog settings, and CO2 and CH4 logging config. |
2017-10-04 | AIU-1570 | LI-7550 | Removed | Interface LI-7550 | Removed from the Bouldin Corn tower for testing in the lab for bad headers and 7700 drop outs during wash cycle. Replaced by sn AIU-0614. |
2017-10-04 | AIU-0614 | LI-7550 | Installed | Interface LI-7550 | Installed on the Bouldin Corn tower to replace sn AIU-1570 for testing in the lab for bad headers and 7700 drop outs during wash cycle. |
2019-12-02 | AIU-0614 | LI-7550 | Removed | Interface LI-7550 | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | AIU-0614 | LI-7550 | Installed | Interface LI-7550 | Re-installed at Bouldin Corn after re-assembling tower. |
2021-01-29 | AIU-0614 | LI-7550 | Removed | Interface LI-7550 | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. |
2021-02-03 | AIU-1866 | LI-7550 | Installed | Interface LI-7550 | Installed on the BC tower to replace sensors damaged when the tower fell last week |
2021-04-15 | AIU-1866 | LI-7550 | Upgrade | Interface LI-7550 | Upgraded firmware from v8.5.0 to v8.8.32 |
2022-05-27 | 13056 | NL115 | Removed | Interface NL115 | I could not get this card reader to work with either of two CF cards. Removed for testing in the lab. |
2022-06-09 | 6761 | NL115 | Installed | Interface NL115 | Installed on met CR1000 at BC and connected to Stardot phenocam |
2022-07-14 | 6761 | NL115 | Removed | Interface NL115 | Removed from BC and brought back to the lab for testing. The bottom 1/2-1/3 of phenocam photos is gray. Could be an interface or a card problem. It was replaced by sn 13056. |
2022-07-14 | 13056 | NL115 | Installed | Interface NL115 | Installed on the BC met CR1000 to replace sn 6761, which was brought back to the lab for testing. |
2023-03-02 | AIU-1866 | LI-7550 | Removed | Interface LI-7550 | Unit is receiving 24V but no LED lights and couldn't connect. Removed from BC tower and brought back to the lab for testing. |
2023-03-08 | AIU-0674 | LI-7550 | Installed | Interface LI-7550 | Installed at BC tower to replace sn AIU-1866 which was removed because of power problems |
Multiplexer | |||||
2017-04-26 | 21690 | AM16/32B | Installed | Multiplexer Met | Installed at the Bouldin Corn site with the initial set up of the tower. |
2017-04-26 | 7908 | AM25T | Installed | Multiplexer Thermocouple | Installed at the Bouldin Corn site with the initial set up of the tower. |
2019-12-02 | 7908 | AM25T | Removed | Multiplexer Thermocouple | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 21690 | AM16/32B | Removed | Multiplexer Met | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | 21690 | AM16/32B | Installed | Multiplexer Met | Re-installed at Bouldin Corn after re-assembling tower. |
2019-12-13 | 7908 | AM25T | Installed | Multiplexer Thermocouple | Re-installed at Bouldin Corn after re-assembling tower. |
Narrow Band | |||||
2017-04-26 | 913003004 | SRS-Pi PRI | Installed | Narrow Band PRI Incoming | Installed at the Bouldin Corn site with the initial set up of the tower. |
2017-04-26 | 882103020 | SRS-Ni NDVI | Installed | Narrow Band NDVI Incoming | Installed at the Bouldin Corn site with the initial set up of the tower. |
2017-04-26 | 943103015 | SRS-Nr NDVI | Installed | Narrow Band NDVI Reflected | Installed at the Bouldin Corn site with the initial set up of the tower. |
2017-04-26 | 897103002 | SRS-Pi PRI | Installed | Narrow Band PRI Reflected | Installed on the Bouldin Corn Tower with the initial set up of the site. This sensor is in the reflected position using SDI-12 address F |
2017-06-05 | 943103015 | SRS-Nr NDVI | Removed | Narrow Band NDVI Reflected | Removed from BC tower and replaced by a downward-facing NDVI sensor with hemispherical view. |
2017-06-05 | 2046503357 | SRS-Ni NDVI | Installed | Narrow Band NDVI Reflected | Installed at BC tower to replace downward-facing NDVI sensor with field-stop lens. |
2019-12-02 | 882103020 | SRS-Ni NDVI | Removed | Narrow Band NDVI Incoming | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 913003004 | SRS-Pi PRI | Removed | Narrow Band PRI Incoming | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 897103002 | SRS-Pi PRI | Removed | Narrow Band PRI Reflected | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 2046503357 | SRS-Ni NDVI | Removed | Narrow Band NDVI Reflected | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | 882103020 | SRS-Ni NDVI | Installed | Narrow Band NDVI Incoming | Re-installed at Bouldin Corn in the incoming position after re-assembling tower. |
2019-12-13 | 2046503357 | SRS-Ni NDVI | Installed | Narrow Band NDVI Reflected | Re-installed at Bouldin Corn in the outgoing position after re-assembling tower. |
2019-12-13 | 913003004 | SRS-Pi PRI | Installed | Narrow Band PRI Incoming | Re-installed at Bouldin Corn after re-assembling tower. This hemispherical PRI sensor was installed in the incoming position. |
2019-12-13 | 930103952 | SRS-Pr PRI | Installed | Narrow Band PRI Reflected | Re-installed at Bouldin Corn after re-assembling tower. This field-stop PRI sensor is in the outgoing position. |
2020-01-07 | 930103952 | SRS-Pr PRI | Removed | Narrow Band PRI Reflected | Removed from Bouldin Corn. This is a field stop (narrow FOV) sensor and we prefer the hemispherical FOV sensors. This one was put up because we thought the hemispherical one was not working but it turned out to be a program bug instead. |
2020-01-07 | 897103002 | SRS-Pi PRI | Installed | Narrow Band PRI Reflected | Installed on Bouldin Corn in the outgoing position to replace a field stop (narrow FOV) sensor. We prefer the hemispherical senors. We thought this sensor was not working but it was a program bug not a sensor problem. |
2021-01-29 | 882103020 | SRS-Ni NDVI | Removed | Narrow Band NDVI Incoming | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
2021-01-29 | 2046503357 | SRS-Ni NDVI | Removed | Narrow Band NDVI Reflected | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
2021-01-29 | 913003004 | SRS-Pi PRI | Removed | Narrow Band PRI Incoming | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. |
2021-01-29 | 897103002 | SRS-Pi PRI | Removed | Narrow Band PRI Reflected | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
2021-02-03 | 913003004 | SRS-Pi PRI | Installed | Narrow Band PRI Incoming | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
2021-02-03 | 882103020 | SRS-Ni NDVI | Installed | Narrow Band NDVI Incoming | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
2021-02-03 | 2046503357 | SRS-Ni NDVI | Installed | Narrow Band NDVI Reflected | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
2021-02-03 | 897103002 | SRS-Pi PRI | Installed | Narrow Band PRI Reflected | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
2021-04-15 | 1051 | S2-411 | Installed | Narrow Band NDVI Incoming | Installed on BC tower in incoming position |
2021-04-15 | 1052 | S2-411 | Installed | Narrow Band NDVI Reflected | Installed on BC tower in outgoing position |
2024-02-28 | 882103020 | SRS-Ni NDVI | Removed | Narrow Band NDVI Incoming | Removed from BC tower because most of the SRS NDVI sensors installed in the Delta have been drifting over the past few years |
2024-02-28 | 2046503357 | SRS-Ni NDVI | Removed | Narrow Band NDVI Reflected | Removed from BC tower because most of the SRS NDVI sensors installed in the Delta have been drifting over the past few years |
Net Radiometer | |||||
2017-04-26 | 2420 | NR01 | Installed | Net Radiometer 4-way | Installed at the Bouldin Corn site with the initial set up of the tower. |
2017-05-10 | 990181 | CNR1 | Installed | Net Radiometer 4-way | Installed at Bouldin Corn to calibrate against NR01 sn 2420. Joe had replaced the sensor cables recently. |
2017-05-25 | 990181 | CNR1 | Removed | Net Radiometer 4-way | Removed from BC tower because comparison is complete |
2021-01-29 | 2420 | NR01 | Removed | Net Radiometer 4-way | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged and broken in two pieces |
2021-02-03 | 990181 | CNR1 | Installed | Net Radiometer 4-way | Installed on the BC tower to replace the Hukseflux NR01 that broke when the tower fell and still needs repair and cross comparison. |
PAR Sensor | |||||
2017-04-26 | 140453 | PQS 1 | Installed | PAR Sensor Incoming | Installed in the incoming position at the Bouldin Corn site with the initial set up of the tower. |
2017-04-26 | 140441 | PQS 1 | Installed | PAR Sensor Reflected | Installed in the outgoing position at the Bouldin Corn site with the initial set up of the tower. |
2017-07-14 | 140456 | PQS 1 | Installed | PAR Sensor Diffuse | Installed at Bouldin Corn tower as part of a diffuse PAR sensor |
2019-12-02 | 140456 | PQS 1 | Removed | PAR Sensor Diffuse | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 140453 | PQS 1 | Removed | PAR Sensor Incoming | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 140441 | PQS 1 | Removed | PAR Sensor Reflected | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | 140441 | PQS 1 | Installed | PAR Sensor Reflected | Re-installed at Bouldin Corn in outgoing position after re-assembling tower. |
2019-12-13 | 140453 | PQS 1 | Installed | PAR Sensor Incoming | Re-installed at Bouldin Corn in incoming position after re-assembling tower. |
2019-12-13 | 140456 | PQS 1 | Installed | PAR Sensor Diffuse | Re-installed at Bouldin Corn after re-assembling tower as part of a diffuse PAR sensor. |
2020-05-29 | 040412 | PAR-LITE | Installed | PAR Sensor Under-A | I stalled about 10cm above the ground near the picams to measure under canopy PAR |
2020-05-29 | 030381 | PAR-LITE | Installed | PAR Sensor Under-B | I stalled about 10cm above the ground near the picams to measure under canopy PAR |
2020-08-05 | 030381 | PAR-LITE | Removed | PAR Sensor Under-B | Removed from under the canopy at Bouldin Corn. The corn has reached maximum growth continued measurements of questionable use. |
2020-08-05 | 040412 | PAR-LITE | Removed | PAR Sensor Under-A | Removed from under the canopy at Bouldin Corn. The corn has reached maximum growth continued measurements of questionable use. |
2021-01-29 | 140453 | PQS 1 | Removed | PAR Sensor Incoming | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
2021-01-29 | 140441 | PQS 1 | Removed | PAR Sensor Reflected | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Completely submerged |
2021-02-03 | 140453 | PQS 1 | Installed | PAR Sensor Incoming | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
2021-02-03 | 140441 | PQS 1 | Installed | PAR Sensor Reflected | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
Precipitation | |||||
2017-04-26 | 65548-615 | TE525MM | Installed | Precipitation | Installed on the top southeast corner of the scaffolding at the Bouldin Corn site with the initial set up of the tower. |
2018-10-18 | 65548-615 | TE525MM | Calibration | Precipitation | Got 94 of 100 tips. Adjusted a total of 1 turn CW. 2018-10-18: Got 94 of 100 tips - adjusted screws 0.5 turns CW 2018-10-25: Got 97 tips and adjusted set screws 0.5 turn CW. 2018-11-01: 99 of 100 tips - no adjustment. |
2019-10-22 | 65548-615 | TE525MM | Calibration | Precipitation | After cleaning the tipping bucket we connected it to a CR23x and dripped 475ml of water through the bucket which should result in 100tips. We got 99 tips – no adjustment needed. |
2019-12-02 | 65548-615 | TE525MM | Removed | Precipitation | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | 65548-615 | TE525MM | Tested | Precipitation | Plugged tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket which should result in 100 tips. Got 102 tips and did not adjust the bucket. |
2019-12-13 | 65548-615 | TE525MM | Installed | Precipitation | Re-installed at Bouldin Corn after re-assembling tower. |
2020-11-19 | 65548-615 | TE525MM | Precipitation | Plugged tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket which should result in 100 tips. Got 116 tips. Did not adjust - will retest with better water metering system. | |
2020-12-03 | 65548-615 | TE525MM | Tested | Precipitation | Plugged tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket (1/2 of the bottle at a time) which should result in 100 tips. Got 117 tips. Did not adjust - will retest with better water metering system. |
2021-10-21 | 65548-615 | TE525MM | Tested | Precipitation | We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 54 tips and adjusted its set screws 1.5 turns CCW (1/2 turn of both screws is a 2-3% adjustment). |
2021-11-05 | 65548-615 | TE525MM | Tested | Precipitation | We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 52 tips but forgot to adjust the screws. |
2021-11-18 | 65548-615 | TE525MM | Calibration | Precipitation | I adjusted the screw 1 turn CCW before testing b/c I forgot to adjust it after the previous test. We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 49 tips and I called it good until next year. |
Range Finder | |||||
2019-06-20 | URM06-001 | URM06 | Moved | Range Finder | Moved from BA to BC to look at corn height. It was installed on the BC radiation boom. |
2019-12-02 | URM06-001 | URM06 | Removed | Range Finder | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
RH & Temp | |||||
2017-04-26 | L1740395 | HMP60 | Installed | RH & Temp | Installed on the 7500 post at the Bouldin Corn site with the initial set up of the tower. |
2019-12-02 | L1740395 | HMP60 | Removed | RH & Temp | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | L1740395 | HMP60 | Installed | RH & Temp | Re-installed at Bouldin Corn after re-assembling tower. |
2020-07-21 | L1740395 | HMP60 | Removed | RH & Temp | Removed from BC and brought back to the lab for inspection. AirT/RH both went bad on 7/5 but nothing obviously wrong with it. |
2020-07-30 | L1740398 | HMP60 | Installed | RH & Temp | Installed at BC to replace sn L1740395, which was sent to Campbell for repair |
2023-04-03 | L1740398 | HMP60 | Removed | RH & Temp | Removed from the BC tower and brought back to the lab for testing b/c the sensor was not giving any data |
2023-04-11 | E3514 | HygroVUE10 | Installed | RH & Temp | Installed at BC tower at the same height/location to replace RH/T sensor HMP60 sn L1740398. |
Soil Moisture Probe | |||||
2017-04-26 | M003370 | ML3 | Installed | Soil Moisture Probe Theta | Installed at the Bouldin Corn site with the initial set up of the tower. Temporarily at the south edge of the scaffolding until tractor work is finished. |
2017-04-26 | 17129 | CS655 | Installed | Soil Moisture Probe A | Installed at the Bouldin Corn site with the initial set up of the tower. Temporarily at the south edge of the scaffolding until tractor work is finished. |
2017-04-26 | 17128 | CS655 | Installed | Multiplexer Thermocouple | Installed at the Bouldin Corn site with the initial set up of the tower. Temporarily at the south edge of the scaffolding until tractor work is finished. |
2017-04-26 | 17133 | CS655 | Installed | Soil Moisture Probe C | Installed at the Bouldin Corn site with the initial set up of the tower. Temporarily at the south edge of the scaffolding until tractor work is finished. |
2019-12-02 | 17128 | CS655 | Removed | Soil Moisture Probe B | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 17129 | CS655 | Removed | Soil Moisture Probe A | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | 17133 | CS655 | Removed | Soil Moisture Probe C | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | M003370 | ML3 | Removed | Soil Moisture Probe Theta | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | M003370 | ML3 | Installed | Soil Moisture Probe Theta | Re-installed at Bouldin Corn after re-assembling tower. |
2019-12-13 | 17133 | CS655 | Installed | Soil Moisture Probe C | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
2019-12-13 | 17129 | CS655 | Installed | Soil Moisture Probe A | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
2019-12-13 | 17128 | CS655 | Installed | Soil Moisture Probe B | Re-installed at Bouldin Corn after re-assembling tower. Need to check if this in A/B/C location. |
2020-05-19 | M003370 | ML3 | Moved | Soil Moisture Probe Theta | Moved back into the field from being under the scaffolding at the side of the field for tilling and planting. |
2020-05-19 | 17133 | CS655 | Moved | Soil Moisture Probe C | Moved back into the field from being under the scaffolding at the side of the field for tilling and planting. |
2020-05-19 | 17128 | CS655 | Moved | Soil Moisture Probe B | Moved back into the field from being under the scaffolding at the side of the field for tilling and planting. |
2020-05-19 | 17129 | CS655 | Moved | Soil Moisture Probe A | Moved back into the field from being under the scaffolding at the side of the field for tilling and planting. |
2020-09-01 | M003370 | ML3 | Moved | Soil Moisture Probe Theta | Moved tower and soil sensors back from field in preparation for corn harvest. The theta probe was reinstalled vertically under the tower. |
2020-09-01 | 17133 | CS655 | Moved | Soil Moisture Probe C | Moved tower and soil sensors back from field in preparation for corn harvest. Soil sensors were reinstalled under the tower at -2cm depth. |
2020-09-01 | 17128 | CS655 | Moved | Soil Moisture Probe B | Moved tower and soil sensors back from field in preparation for corn harvest. Soil sensors were reinstalled under the tower at -2cm depth. |
2020-09-01 | 17129 | CS655 | Moved | Soil Moisture Probe A | Moved tower and soil sensors back from field in preparation for corn harvest. Soil sensors were reinstalled under the tower at -2cm depth. |
2022-03-31 | 17129 | CS655 | Moved | Soil Moisture Probe A | Moved from the field back under the tower in anticipation of tilling. This sensor is to the north at 10cm depth. |
2022-03-31 | 17133 | CS655 | Moved | Soil Moisture Probe C | Moved from the field back under the tower in anticipation of tilling. This sensor is horizontally between the other 2 Campbell sensors at 20cm depth. |
2022-03-31 | 17128 | CS655 | Moved | Soil Moisture Probe B | Moved from the field back under the tower in anticipation of tilling. This sensor is to the south at 15cm depth. |
2022-03-31 | M003370 | ML3 | Moved | Soil Moisture Probe Theta | Moved from the field back under the tower in anticipation of tilling. This sensor is buried just under the surface. |
2022-05-27 | 17128 | CS655 | Moved | Soil Moisture Probe B | Moved from under the tower in to the field after planting southwest of the tower. The tower was moved about 2m west to the edge of the field |
2022-05-27 | 17133 | CS655 | Moved | Soil Moisture Probe C | Moved from under the tower in to the field after planting west of the tower. The tower was moved about 2m west to the edge of the field |
2022-05-27 | 17129 | CS655 | Moved | Soil Moisture Probe A | Moved from under the tower in to the field after planting northwest of the tower. The tower was moved about 2m west to the edge of the field |
2023-04-26 | 17128 | CS655 | Moved | Soil Moisture Probe B | Moved from field to tower in preparation for impending plowing. Not installed in soil yet. |
2023-04-26 | 17129 | CS655 | Moved | Soil Moisture Probe A | Moved from field to tower in preparation for impending plowing. Not installed in soil yet. |
2023-04-26 | 17133 | CS655 | Moved | Soil Moisture Probe C | Moved from field to tower in preparation for impending plowing. Not installed in soil yet. |
2023-05-03 | 17128 | CS655 | Installed | Soil Moisture Probe B | Installed in soil under tower. Probe was installed horizontally with top of probe 5cm under soil surface. |
2023-05-03 | 17129 | CS655 | Installed | Soil Moisture Probe A | Installed in soil under tower. Probe was installed horizontally with top of probe 5cm under soil surface. |
2023-05-03 | 17133 | CS655 | Installed | Soil Moisture Probe C | Installed in soil under tower. Probe was installed horizontally with top of probe 5cm under soil surface. |
Solar Controller | |||||
2021-06-16 | 2963 | CH150 | Installed | Solar Controller Surface Renewal | Installed as part of Kevin Tu's surface renewal experiment but not being used due to concerns about pulling too much power. |
2024-01-04 | 2963 | CH150 | Removed | Solar Controller Surface Renewal | Removed from BC at the conclusion of Corteva's IRT experiment. |
Water Level | |||||
2019-04-11 | 70011234 | CS451 | Installed | Water Level | Installed in a well underneath the BC tower. Sensor is 92cm below ground level. From field notes: Top of pipe to ground = 25cm Top of pipe to sensor = 117cm Top of pipe to water = 85cm Sensor reading = 0.46 psi ~ 32cm (matches manual reading) |
2019-12-02 | 70011234 | CS451 | Removed | Water Level | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | 70011234 | CS451 | Installed | Water Level | Re-installed at Bouldin Corn after re-assembling tower. Sensor is 91cm below ground level. Top of well to ground = 47 cm Top of well to sensor = 138 cm Top of well to water level = 131 cm (4.3’) Sensor psi reading = 0.12 psi ~ 8 cm (consistent with physical measurement) |
2020-05-19 | 70011234 | CS451 | Moved | Water Level | Dug new well after moving tower into corn field. Sensor is 91cm below ground level. Water level from top of well = 131 cm (4.3’) Top of well to ground = 47 cm Top of well to sensor = 138 cm Sensor psi reading = 0.12 psi ~ 8 cm (checks out with physical measurement!) |
2020-08-05 | 70011234 | CS451 | Notes | Water Level | Measured with a tape measure Top of well to ground = 47 cm Top of well to water surface = 69 cm Water level reading = 0.99 psi ~ 70 cm |
2020-09-01 | 70011234 | CS451 | Moved | none | Moved tower and soil sensors back from field in preparation for corn harvest. Joe dug a new well east of the tower for the water level sensor. Sensor is 69 cm below ground. Top of pipe to ground = 58 cm Top of pipe to sensor = 127 cm Well depth ~ 160 cm (dry at bottom) |
2023-08-24 | 70011234 | CS451 | Removed | Water Level | Removed from Bouldin Corn because its cable is short. Replaced by sn 20010770. |
2023-08-24 | 20010770 | CS450 | Installed | Water Level | Installed at Bouldin Corn to replace sn 70011234 for longer cable. |
Weather Station | |||||
2018-06-14 | A000251 | Mark I | Installed | Weather Station A | Installed on the upper railing between the eddy and radiation sensors. It is pointed west looking at the corn field. |
2018-10-04 | A000251 | Mark I | Removed | Weather Station A | Removed from Bouldin Corn to be sent back to Arable for inspection of back up temperature sensor. Replaced by sn A001911. Later we realized we removed the incorrect Arable and they had requested instead sn A000591. |
2018-10-04 | A001911 | Mark I | Installed | Weather Station A | Installed on Bouldin Corn to replace sn A000251 which is to be sent back to Arable for inspection of backup temperature sensor. Later we realized we had sent back the incorrect Arable; they had requested sn A000591. |
2019-04-25 | A001359 | Mark I | Installed | Weather Station B | Installed by Loreli and Mikaela from Arable for inter-comparison experiment. |
2019-05-23 | C002445 | Mark II | Installed | Weather Station C | Installed on a new boom fixed to the southwest corner of the scaffolding for inter-comparison with existing Mark I unit and our tower data. The Mark I unit was also moved to the new boom. |
2019-06-20 | A001911 | Mark I | Removed | Weather Station A | Removed from BC and brought back to the lab. There's nothing wrong with it, it was just the oldest of 3 Arables on the BC tower. |
2019-12-02 | C002445 | Mark II | Removed | Weather Station C | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-02 | A001359 | Mark I | Removed | Weather Station B | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2019-12-13 | A001359 | Mark I | Installed | Weather Station B | Re-installed at Bouldin Corn after re-assembling tower. |
2020-04-17 | C003251 | Mark II | Installed | Weather Station C | Installed at BC to replaced the prototype unit that was disabled and removed when the tower fell over. |
2020-06-04 | C003771 | Mark II | Installed | Weather Station PL_A | Installed on a telescoping post in normal corn northeast of the tower. This is part of the Planet Labs / Arable inter-comparison project. |
2020-06-04 | C004318 | Mark II | Installed | Weather Station PL_B | Installed on a telescoping post in normal corn northeast of the tower. This is part of the Planet Labs / Arable inter-comparison project. |
2021-01-21 | A001359 | Mark I | Removed | Weather Station B | Removed from Bouldin Corn as Mark I's are phased out. Replaced by Mark II sn C006123. |
2021-01-21 | C006123 | Mark II | Installed | Weather Station B | Installed at Bouldin Corn to replace Mark I sn A001359 which is being phased out. |
2021-01-29 | C003251 | Mark II | Removed | Weather Station C | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Maybe some rain in the Tair bulge |
2021-01-29 | C006123 | Mark II | Removed | Weather Station B | Removed due to the tower falling over in the storm near midnight on 2021-01-26/27. Maybe some rain in the Tair bulge |
2021-02-03 | C003251 | Mark II | Installed | Weather Station C | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
2021-02-03 | C006123 | Mark II | Installed | Weather Station B | Reinstalled on the BC tower after cleaning and inspection in the lab after being removed after the tower fell last week |
2021-06-16 | C003771 | Mark II | Removed | Weather Station PL_A | Removed per request from Arable |
2022-03-31 | C004318 | Mark II | Removed | Weather Station PL_B | Removed from Bouldin Corn and replaced by sn C011081. |
2022-03-31 | C011081 | Mark II | Installed | Weather Station PL_B | Installed at Bouldin Corn to replace sn C004318. |
2022-11-18 | B301069 | Mark 3 | Installed | none | Installed production Mark 3 on the west side of the scaffolding. This replaced a prototype Mark 3 that we removed today. |
2022-11-30 | C003251 | Mark II | Removed | Weather Station C | Removed and brought back to the lab because we don't need 3 Arable units on one tower. We added a Mark 3 last visit and there is a second Mark II on the tower. |
2023-09-12 | C011081 | Mark II | Removed | Weather Station PL_B | Removed with post and ground anchor for a location near the road to the southeast of the tower. The PL-Arable project is paused according to PL. All for status lights were green when I shut it down. |
2024-02-14 | B301069 | Mark 3 | Removed | Removed from Bouldin Corn and replaced with the updated Mark 3 sn D006617. This older Mark 3 will be sent back to Loreli. | |
2024-02-14 | D006617 | Mark 3 | Installed | none | Installed at Bouldin Corn with a new 2-D sonic, sn 022365410, to replace the older Mark 3, sn B301069. |