View history entry 4294 for Campbell Sci. CS451 sn: 70011234

Date: 2019-12-13
From: Lab - none
To: Bouldin Corn - Water Level
Title: Installed
Notes: Re-installed at Bouldin Corn after re-assembling tower. Sensor is 91cm below ground level. Top of well to ground = 47 cm Top of well to sensor = 138 cm Top of well to water level = 131 cm (4.3’) Sensor psi reading = 0.12 psi ~ 8 cm (consistent with physical measurement)
Entered by: Daphne Szutu