View history entry 3924 for Campbell Sci. CS451 sn: 70011234

Date: 2019-04-11
From: Lab - none
To: Bouldin Corn - Water Level
Title: Installed
Notes: Installed in a well underneath the BC tower. Sensor is 92cm below ground level. From field notes: Top of pipe to ground = 25cm Top of pipe to sensor = 117cm Top of pipe to water = 85cm Sensor reading = 0.46 psi ~ 32cm (matches manual reading)
Entered by: Daphne Szutu