Equipment and Calibration History
History for YSI EXO fDOM 22C100780
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6943 | 2024-10-23 | Lab none | Installed | Gilbert Tract fDOM | Installed in the GT channel after calibration |
7038 | 2024-10-22 | Lab none | Calibration | Lab none | fDOM (QSU and RFU), Last cal: EXO sn 23H107001 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-1.75 QSU,0.00 QSU,0.00 QSU,23.325 °C Pt2,300.0 QSU,272.10 QSU,300.00 QSU,0.00 QSU,22.418 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,-0.61 RFU,0.00 RFU,0.00 RFU,23.221 °C Pt2,100.00 RFU,91.28 RFU,100.00 RFU,0.00 RFU,22.440 °C Note: Made fresh calibration solution using QSE solution from 2024-07-20. Calibrated with a mess-up calibration solution yesterday so did a factory reset before today's calibration. |
6927 | 2024-10-17 | Gilbert Tract fDOM | Removed | Lab none | Removed from GT channel and brought back to the lab for calibration |
6827 | 2024-08-16 | Lab none | Installed | Gilbert Tract fDOM | Installed with EXO3s in Gilbert Tract channel, co-located with the ADCP and channel miniDOT |
6792 | 2024-07-24 | Lab none | Calibration | Lab none | fDOM (QSU and RFU), Last cal: 10/27/2023 14:54 AM EXO sn 23H107001 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-2.37 QSU,0.00 QSU,0.00 QSU,23.604 °C Pt2,313.80 QSU,290.85 QSU,313.83 QSU,0.00 QSU,12.006 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,-0.84 RFU,0.00 RFU,0.00 RFU,24.327 °C Pt2,104.60 RFU,91.47 RFU,104.61 RFU,0.00 RFU,12.002 °C Note: Used calibration solution from 2024-07-21. This sensor was returned to us from USGS on 2024-07-23 at Gilbert Tract. |
6771 | 2024-07-23 | Gilbert Tract | Removed | Lab none | Returned to us from USGS. Was deployed with their EXO in the Gilbert Tract channel. |
6060 | 2022-11-03 | Gilbert Tract | Calibration | Gilbert Tract | fDOM (QSU and RFU), Last cal: 10/24/2022 8:08:51 AM EXO sn 22C100485 EXO/Sensor Firmware: 1.0.86/3.0.0 UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-1.28 QSU,0.00 QSU,0.00 QSU,16.696 °C Pt2,316.50 QSU,315.87 QSU,316.54 QSU,0.00 QSU,10.444 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,0.11 RFU,0.00 RFU,0.00 RFU,10.610 °C Pt2,105.50 RFU,92.32 RFU,105.51 RFU,0.00 RFU,11.275 °C |
5826 | 2022-05-06 | Lab none | Installed | Gilbert Tract none | Mckenna from USGS installed this at the GT breach site next to the other USGS water sensors in the channel. They plan to service the sensors every 3 months. |