Equipment and Calibration
View History
Equipment Entry
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6264 | 2022-08-24 | Factory none | Disposed | Disposed none | This unit was originally borrowed from Eosense. Since this model is no longer in production they can not repair it ahead of other purchased units so we will not be getting it back. |
6263 | 2022-08-23 | Lab none | Shipped | Factory none | Sent to Eosense on RMA #220809 for repair. |
5579 | 2021-11-17 | Vaira Forced Diffusion West | Removed | Lab none | Removed from Vaira West due to bad CO2 concentration readings |
5373 | 2021-07-13 | Vaira Forced Diffusion West | Maintenance | Vaira Forced Diffusion West | Performance check done. Referenced to 400ppm. Average Flux: -0.0, 0.08 STDEV Average ATM: 610.7ppm, 49.6C Average Soil: 609.8ppm, 49.8C |
5218 | 2021-03-17 | Lab none | Installed | Vaira Forced Diffusion West | Installed at Vaira to replace sn FD20190028 for possible use at Hill Slough. |
5200 | 2021-03-15 | Factory none | Returned | Lab none | Returned from Eosense. It was sent in for zero CO2 readings but Eosense reported that it seemed to be reading correctly. It was calibrated from CO2 concentration and CO2 flux. |
5199 | 2021-01-12 | Lab none | Shipped | Factory none | Sent to Eosense for repair due to zero CO2 readings |
4932 | 2020-10-22 | Tonzi-SoilCO2 Forced Diffusion Tree 1 | Removed | Lab none | Removed from the under tree location at Tonzi because its co2 sensor has been reading zero. Replaced by sn FD20180048 |
4173 | 2019-11-19 | Lab none | Installed | Tonzi Forced Diffusion Tree 1 | Installed under the tree at the Tonzi soil co2 profiles |