Equipment and Calibration History
History for Texas Electronics TR-525I-R1 24872-999
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5478 | 2021-10-06 | Tonzi | Calibration | Tonzi none | Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 54 tips – I turned the screws two turns CCW |
4124 | 2019-10-24 | Tonzi | Calibration | Tonzi | I connected the bucket to a CR23x dripping 475ml of water through the bucket which should result in 100tips. I got 100 tips. |
3736 | 2018-10-17 | Tonzi | Calibration | Tonzi | Got 88 of 100 tips - adjusted screws 2.5 turns CW |