Report: Tonzi Floor well sensor comparison - 2018-03-27
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5256 | 2021-05-04 | Lab none | Installed | Tonzi Water Level Floor | Reinstalled after cows chewed the extension cable and Joe couldn't figure out the wiring. |
5252 | 2021-04-29 | Lab none | Tested | Lab none | I couldn't get this sensor to work in the field. Seem to test okay on a CR1000 terminal using the SDI-12 pass through. Will try again in the field again next time. It's using address 0 (zero): 0I! - id, 0M! - Measure, 0D0! - Data. Note CRBasic program automatically issues 0D0! command after 0M! command. |
5246 | 2021-04-18 | Tonzi Water Level Floor | Removed | Lab none | I found it's cable chewed. I cut of the chewed part and reattached but still no data. Power and data lines seems continuous. Maybe chewing shorted out something. |
3324 | 2018-02-07 | Lab none | Installed | Tonzi | Installed in the well near the Floor tower. The sensor is 11.1m below the top of the white pipe. It is connected to the CR200. |
3328 | 2018-02-06 | Lab none | Maintenance | Lab none | Extended the cable and swapped data and power colors so that white is wired to the data port and red is wired to the 12V port. Decagon colors --> Extension colors --> datalogger Red --> White --> Data White --> Red --> 12V Black --> Black --> G Bare --> Green --> AG |