Equipment and Calibration History

History for Decagon-METER CTD-10 1260403156

Rec No.DateLoc FromTitleLoc ToNotes
5256 2021-05-04 Lab none Installed Tonzi Water Level Floor Reinstalled after cows chewed the extension cable and Joe couldn't figure out the wiring.
5252 2021-04-29 Lab none Tested Lab none I couldn't get this sensor to work in the field. Seem to test okay on a CR1000 terminal using the SDI-12 pass through. Will try again in the field again next time. It's using address 0 (zero): 0I! - id, 0M! - Measure, 0D0! - Data. Note CRBasic program automatically issues 0D0! command after 0M! command.
5246 2021-04-18 Tonzi Water Level Floor Removed Lab none I found it's cable chewed. I cut of the chewed part and reattached but still no data. Power and data lines seems continuous. Maybe chewing shorted out something.
3324 2018-02-07 Lab none Installed Tonzi Installed in the well near the Floor tower. The sensor is 11.1m below the top of the white pipe. It is connected to the CR200.
3328 2018-02-06 Lab none Maintenance Lab none Extended the cable and swapped data and power colors so that white is wired to the data port and red is wired to the 12V port.
Decagon colors --> Extension colors --> datalogger
Red --> White --> Data
White --> Red --> 12V
Black --> Black --> G
Bare --> Green --> AG