Report: Mayberry conductivity comparision - 2020 - 2020-11-03
Report: A547 board calibration - 2020 - 2020-12-08
Report: Mayberry porewater sippers - 2023-05-09
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5032 | 2020-12-09 | Lab none | Installed | Mayberry Conductivity Sensor A | Installed at MB with interface sn 547-02 to replace sn 5688, which was brought back to the lab for testing. |
4980 | 2020-11-19 | Mayberry | Removed | Lab none | Removed sensor and factory A547 interface from MB after concluding the conductivity sensor comparison. Brought back to the lab to test a homemade A547 board. |
4943 | 2020-11-03 | Sherman Wetland Conductivity Sensor | Moved | Mayberry none | Moved from SW to MB for a temporary intercomparison because there was no water at the SW tower and sn 5688 at MB was giving high temp number |
4499 | 2020-05-26 | Sherman Wetland Conductivity Sensor | Calibration | Sherman Wetland Conductivity Sensor | Cleaned and calibrated in the field with DI water (0 mS) and a YSI solution (10,000 uS). These are the temperature-corrected mScm readings from the datalogger, where the temperature correction occurs in the datalogger program.
In DI water: 0.56 mScm In 10,000 uS calibration soln: 10.6 mScm Calibration: (Corrected mScm) = 0.996 * (Sensor mScm) - 0.558 |
2646 | 2016-05-24 | Lab none | Installed | Sherman Wetland Conductivity Sensor | Installed at Sherman Wetland |