Equipment and Calibration

View History

Equipment Entry

Instrument:Conductivity Sensor
Manufacture:Campbell Sci.
Model No:CS547A
Serial No:6494
Tag No:NULL
Acquisition Date:2013-12-18
Project:DWR-Wetland
Location on 2020-05-29:East End, NULL
Notes: Kc: 1.418
Lead length: 20'

Documents

No documents found

Report: A547 board calibration - 2014 - 2014-06-09

History

Rec No.DateLoc FromTitleLoc ToNotes
4508 2020-05-29 East End Conductivity Sensor Calibration East End Conductivity Sensor Calibration in the field with DI water (0 mS) and a YSI solution (10,000 uS). Temperature-corrected mS/cm readings from the data logger, where the temperature correction occurs in the data logger program.
In DI water: 0.00mS/cm, 22.4C
In 10,000 uS calibration soln: 11.3mS/cm, 23.0C
Calibration: (Corrected mScm) = 0.8850 * (Sensor mScm)
4310 2019-12-17 Lab none Installed East End Conductivity Sensor Reinstalled at East End after being removed after getting wet ans possibly causing a short circuit. Tested fine in the lab.
4274 2019-12-05 East End Conductivity Sensor Removed Lab none Removed from EE and brought back to the lab for drying and testing. The datalogger door was open when we arrived and the conductivity sensor interface was wet.
4273 2017-09-26 Lab none Installed East End Conductivity Sensor Installed at EE, after I'm assuming it tested fine in the lab.
3073 2017-09-13 East End Conductivity Sensor Removed Lab none Removed from EE because the values were flat.
2382 2015-01-20 Lab none Installed East End Conductivity Sensor Reinstalled after being removed when the scaffolding fell over.
2381 2015-01-05 East End Conductivity Sensor Removed Lab none Removed from the site due to the tower falling over in the wind on 2014-12-30. Likely undamaged.
2376 2014-06-12 Lab none Installed East End Conductivity Sensor Installed on the East End tower with interface sn 3585.