History for Campbell Sci. CS547A 6494
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
4508 | 2020-05-29 | East End Conductivity Sensor | Calibration | East End Conductivity Sensor | Calibration in the field with DI water (0 mS) and a YSI solution (10,000 uS). Temperature-corrected mS/cm readings from the data logger, where the temperature correction occurs in the data logger program. In DI water: 0.00mS/cm, 22.4C In 10,000 uS calibration soln: 11.3mS/cm, 23.0C Calibration: (Corrected mScm) = 0.8850 * (Sensor mScm) |
4310 | 2019-12-17 | Lab none | Installed | East End Conductivity Sensor | Reinstalled at East End after being removed after getting wet ans possibly causing a short circuit. Tested fine in the lab. |
4274 | 2019-12-05 | East End Conductivity Sensor | Removed | Lab none | Removed from EE and brought back to the lab for drying and testing. The datalogger door was open when we arrived and the conductivity sensor interface was wet. |
4273 | 2017-09-26 | Lab none | Installed | East End Conductivity Sensor | Installed at EE, after I'm assuming it tested fine in the lab. |
3073 | 2017-09-13 | East End Conductivity Sensor | Removed | Lab none | Removed from EE because the values were flat. |
2382 | 2015-01-20 | Lab none | Installed | East End Conductivity Sensor | Reinstalled after being removed when the scaffolding fell over. |
2381 | 2015-01-05 | East End Conductivity Sensor | Removed | Lab none | Removed from the site due to the tower falling over in the wind on 2014-12-30. Likely undamaged. |
2376 | 2014-06-12 | Lab none | Installed | East End Conductivity Sensor | Installed on the East End tower with interface sn 3585. |