Equipment and Calibration
View History
Equipment Entry
|
Documents
Calibration: cal_KippZonen_PQS1_sn140455_20140909.pdfPackingSlip: ps_KippZonen_20151021.pdf
Report: PAR comparison 09-2018 - 2018-09-28
Report: Oxford Tract Radiometer Comparison - 06/2019 - 2019-06-18
Report: Vaira PAR Cross Compare - 2020-06-12
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5668 | 2022-01-20 | East End PAR Sensor Undercanopy | Moved | East End PAR Sensor Diffuse | Moved locations at EE from undercanopy PAR 1 to dfPAR on top of the tower |
5229 | 2021-03-30 | Lab none | Installed | East End none | Installed in EE wetland to measure undercanopy PAR. This sensor is SE of the tower mounted on a horizontal boom. The sensor is sitting on top of a fallen bundle of dried reeds. |
4709 | 2020-08-26 | Vaira PAR Sensor Reflected Cal | Removed | Lab none | Removed from Vaira reflected after inter-comparison. |
4646 | 2020-07-28 | Vaira PAR Sensor Incoming Cal | Moved | Vaira PAR Sensor Reflected Cal | Moved from the incoming position to the reflected position to continue inter-comparison |
4640 | 2020-06-12 | Lab none | Installed | Vaira PAR Sensor Incoming Cal | Installed on the Vaira Tower mounted in the incoming position for cross comparison and eventual swap of the old sensor on the tower for cable repair. |
4075 | 2019-08-27 | Lab none | Calibration | Lab none | New calibration factor of 8.30 uV/(umol/m2/s). Tested this sensor at Oxford Tract against AMP reference sensor K&Z PQS1 sn 161073. See RT Report "Oxford Tract Radiometer Comparison - 06/2019" for details. |
4534 | 2018-09-28 | Lab none | Calibration | Lab none | Calibrated in the lab by sticking it out the south-facing window in Hilgard 205 with the reference sensor sn 161073. The factory calibration was 8.15 µV/(µmol/m2·s) and the new calibration was 8.47 µV/(µmol/m2·s). For details see the RT report "PAR comparison 09-2018" https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=14 |
1981 | 2014-09-09 | Factory | Calibration | Factory | Initial factory calibration: 8.15 uV/(umol/m2/s), 470 Ohm |