Rec No. | Date | Loc From | Title | Loc To | Notes |
4656 |
2020-08-04 |
Factory none |
Disposed |
Disposed none |
The sensor was unrepairable and Campbell Sci disposed of it. Note from Campbell: "Sensor is shorted on the 12V input. Sensor is completely potted except for RH and Temperature components. Sensor cannot be repaired." |
4647 |
2020-07-28 |
Lab none |
Shipped |
Factory none |
Sent back to Campbell for repair on RMA # 32280. |
4629 |
2020-07-23 |
Lab none |
Tested |
Lab none |
Tested in the lab. I used a cable in the lab to power the sensor and measure its temperature signal with a volt meter. The reading was wrong and the base of the sensor definitely got warm. I same cable with another HMP60 and got a much more reasonable temperature reading and did not notice the sensor getting warm. The sensor needs to go back to Campbell for evaluation. |
4627 |
2020-07-21 |
Bouldin Corn RH & Temp |
Removed |
Lab none |
Removed from BC and brought back to the lab for inspection. AirT/RH both went bad on 7/5 but nothing obviously wrong with it. |
4297 |
2019-12-13 |
Lab none |
Installed |
Bouldin Corn RH & Temp |
Re-installed at Bouldin Corn after re-assembling tower. |
4265 |
2019-12-02 |
Bouldin Corn RH & Temp |
Removed |
Lab none |
Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2724 |
2017-04-26 |
Lab none |
Installed |
Bouldin Corn RH & Temp |
Installed on the 7500 post at the Bouldin Corn site with the initial set up of the tower. |