Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
4656 | 2020-08-04 | Factory none | Disposed | Disposed none | The sensor was unrepairable and Campbell Sci disposed of it. Note from Campbell: "Sensor is shorted on the 12V input. Sensor is completely potted except for RH and Temperature components. Sensor cannot be repaired." |
4647 | 2020-07-28 | Lab none | Shipped | Factory none | Sent back to Campbell for repair on RMA # 32280. |
4629 | 2020-07-23 | Lab none | Tested | Lab none | Tested in the lab. I used a cable in the lab to power the sensor and measure its temperature signal with a volt meter. The reading was wrong and the base of the sensor definitely got warm. I same cable with another HMP60 and got a much more reasonable temperature reading and did not notice the sensor getting warm. The sensor needs to go back to Campbell for evaluation. |
4627 | 2020-07-21 | Bouldin Corn RH & Temp | Removed | Lab none | Removed from BC and brought back to the lab for inspection. AirT/RH both went bad on 7/5 but nothing obviously wrong with it. |
4297 | 2019-12-13 | Lab none | Installed | Bouldin Corn RH & Temp | Re-installed at Bouldin Corn after re-assembling tower. |
4265 | 2019-12-02 | Bouldin Corn RH & Temp | Removed | Lab none | Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2724 | 2017-04-26 | Lab none | Installed | Bouldin Corn RH & Temp | Installed on the 7500 post at the Bouldin Corn site with the initial set up of the tower. |