Equipment and Calibration History

History for Vaisala HMP60 L1740395

Rec No.DateLoc FromTitleLoc ToNotes
4656 2020-08-04 Factory none Disposed Disposed none The sensor was unrepairable and Campbell Sci disposed of it. Note from Campbell: "Sensor is shorted on the 12V input. Sensor is completely potted except for RH and Temperature components. Sensor cannot be repaired."
4647 2020-07-28 Lab none Shipped Factory none Sent back to Campbell for repair on RMA # 32280.
4629 2020-07-23 Lab none Tested Lab none Tested in the lab. I used a cable in the lab to power the sensor and measure its temperature signal with a volt meter. The reading was wrong and the base of the sensor definitely got warm. I same cable with another HMP60 and got a much more reasonable temperature reading and did not notice the sensor getting warm. The sensor needs to go back to Campbell for evaluation.
4627 2020-07-21 Bouldin Corn RH & Temp Removed Lab none Removed from BC and brought back to the lab for inspection. AirT/RH both went bad on 7/5 but nothing obviously wrong with it.
4297 2019-12-13 Lab none Installed Bouldin Corn RH & Temp Re-installed at Bouldin Corn after re-assembling tower.
4265 2019-12-02 Bouldin Corn RH & Temp Removed Lab none Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26.
2724 2017-04-26 Lab none Installed Bouldin Corn RH & Temp Installed on the 7500 post at the Bouldin Corn site with the initial set up of the tower.