Equipment and Calibration
View History
Equipment Entry
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
4596 | 2020-07-09 | Loaned | Disposed | Disposed | Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus. |
4566 | 2020-07-09 | Sherman Barn Anemometer | Removed | Loaned none | Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus. |
3577 | 2018-07-11 | Lab none | Installed | Sherman Barn Anemometer | Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site. |
3531 | 2018-06-13 | Twitchell Alfalfa Anemometer | Removed | Lab none | Removed with Twitchell Alfalfa site decommissioning. |
2571 | 2017-01-05 | Lab none | Installed | Twitchell Alfalfa Anemometer | Installed at Twitchell Alfalfa to replace sn 153805. Brand new. |
2565 | 2017-01-03 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug.
Old SN: Y153802 New SN: X153802 Upgrade file: Y153802-Y153806-m5.gcu Old firmware: 2329-601 New firmware: 2329-701-01 All Checksum fields report PASS. Log file: WMUpgradeLog_Y153802_170103_110126.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Baud Rate : 19200 Baud Rate : 57600 Output Message Format : M2 Polar Output Message Format : M1 UV(W) Resolution settings : Normal resolution Resolution settings : High resolution Output rate : P1 1 Hz Output rate : P8 20 Hz SOS and Sonic Temperature display : Neither SOS and Sonic Temperature display : SOS Analogue output mode : 0v to 5V Analogue output mode : -5V to 5V Output Channel 1 : Polar angle (wrap mode = 360 degree wrap) Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : Polar Magnitude (full range = 5m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 4 : Status (status mode = Good signal high, bad signal low) Output Channel 4 : Speed of Sound THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER ======================== Serial Number : X153802 ---- SW VERSION -------------------------------------------- Anemometer Firmware Version: 2329-701 ---- HW DETAILS OF ANEMOMETER ------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER ---------------------- Baud Rate : 19200 Baud Rate : 57600 Output Message Format : M2 Polar Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : Normal resolution Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : Neither SOS and Sonic Temperature display : SOS Analogue output mode : 0v to 5V Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : Polar angle (wrap mode = 360 degree wrap) Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : Polar Magnitude (full range = 5m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Status (status mode = Good signal high, bad signal low) Output Channel 4 : Speed of Sound Synchronised Polling : Off |