Equipment and Calibration History

History for Gill WM 1590 153802

Rec No.DateLoc FromTitleLoc ToNotes
4596 2020-07-09 Loaned Disposed Disposed Given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
4566 2020-07-09 Sherman Barn Anemometer Removed Loaned none Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
3577 2018-07-11 Lab none Installed Sherman Barn Anemometer Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
3531 2018-06-13 Twitchell Alfalfa Anemometer Removed Lab none Removed with Twitchell Alfalfa site decommissioning.
2571 2017-01-05 Lab none Installed Twitchell Alfalfa Anemometer Installed at Twitchell Alfalfa to replace sn 153805. Brand new.
2565 2017-01-03 Lab none Upgrade Lab none Upgraded firmware to correct for W bug.

Old SN: Y153802
New SN: X153802
Upgrade file: Y153802-Y153806-m5.gcu
Old firmware: 2329-601
New firmware: 2329-701-01

All Checksum fields report PASS.

Log file: WMUpgradeLog_Y153802_170103_110126.txt

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

-------

THE FOLLOWING FIELDS HAVE CHANGED

Baud Rate : 19200
Baud Rate : 57600
Output Message Format : M2 Polar
Output Message Format : M1 UV(W)
Resolution settings : Normal resolution
Resolution settings : High resolution
Output rate : P1 1 Hz
Output rate : P8 20 Hz
SOS and Sonic Temperature display : Neither
SOS and Sonic Temperature display : SOS
Analogue output mode : 0v to 5V
Analogue output mode : -5V to 5V
Output Channel 1 : Polar angle (wrap mode = 360 degree wrap)
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : Polar Magnitude (full range = 5m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 4 : Status (status mode = Good signal high, bad signal low)
Output Channel 4 : Speed of Sound

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER ========================

Serial Number : X153802

---- SW VERSION --------------------------------------------
Anemometer Firmware Version: 2329-701

---- HW DETAILS OF ANEMOMETER ------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER ----------------------
Baud Rate : 19200
Baud Rate : 57600
Output Message Format : M2 Polar
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : Normal resolution
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : Neither
SOS and Sonic Temperature display : SOS
Analogue output mode : 0v to 5V
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : Polar angle (wrap mode = 360 degree wrap)
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : Polar Magnitude (full range = 5m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Status (status mode = Good signal high, bad signal low)
Output Channel 4 : Speed of Sound
Synchronised Polling : Off