Rec No. | Date | Loc From | Title | Loc To | Notes |
6451 |
2023-10-27 |
Lab none |
Installed |
Gilbert Tract Interface LI-7550 |
Installed with wood screws to 2x4 to replace the broken box removed yesterday |
6215 |
2023-03-14 |
Lab none |
Tested |
Lab none |
I tested this box in the lab. It had been removed from the field because the door was left open and it got wet inside. I had cleaned it up earlier but I'm not sure if I ran it before. Today, it was running but very slowly/laggy. At first I was concerned it was the new laptop but it was the same on the old laptop. I pulled the shroud off and found a little corrosion on one of the ribbon cable connectors. I cleaned this off and it seems to be working ok. |
6143 |
2023-01-03 |
Gilbert Tract Interface LI-7550 |
Removed |
Lab none |
Removed from Gilbert Tract because it was acting funny (took a long time to eject/mount USB, consistent 7700 time sync error) and there was some corrosion inside from being left open last month. it was replaced with AMP box AIU-1039. |
5491 |
2021-09-02 |
Sherman Wetland Gas Analyzer CO2/H2O |
Moved |
Gilbert Tract Gas Analyzer CO2/H2O |
Moved from SW to GT with the decomissioning of SW and the beginning of GT |
5286 |
2021-05-26 |
Sherman Wetland Interface LI-7550 |
Upgrade |
Sherman Wetland Interface LI-7550 |
Upgraded firmware from v8.5.0 to v8.8.32 |
3216 |
2017-11-30 |
Sherman Wetland Interface LI-7550 |
Upgrade |
Sherman Wetland Interface LI-7550 |
Downgraded from embedded software version 8.7.5 to version 8.5.0. This should solve the issue of data-header mismatch that we found in version 8.7.5. |
3160 |
2017-07-14 |
Sherman Wetland Interface LI-7550 |
Upgrade |
Sherman Wetland Interface LI-7550 |
Upgraded LI-7550 software from 8.0.0 to 8.7.5. |
2286 |
2016-05-24 |
Lab none |
Installed |
Sherman Wetland Interface LI-7550 |
Installed on the Sherman Wetland tower at the initial set up of the site. |