Equipment and Calibration History

History for Apogee SI-121 1964

Rec No.DateLoc FromTitleLoc ToNotes
6878 2024-06-04 Lab none Installed Tonzi Infrared Temperature Tree Crown Reinstalled at the little tree northeast of the Tower at Tonzi after having been removed to repair massive cable damage from cows. This sensor is pointed at the tree crown.
6694 2024-04-30 Lab none Repair Lab none The cows had chewed the cable very nearly down to the nub. I had to cut some of the strain relief off to get enough conductor to splice. I added a length of cable and then spliced the end on that has a completion resistor for the thermmistor half bridge. So two splices. It compared withing 0.5C with the other IRT in the window - as good as you get without a fancier setup.
6693 2024-04-10 Tonzi Infrared Temperature Grass A Removed Lab none Removed from the IRT tripod at Tonzi where it was pointed at the ground. The cows ate its cable.
1807 2015-05-06 Lab none Installed Tonzi Infrared Temperature Grass A Installed at Tonzi to the east just north of the road. This sensor is point down on a patch of ground cleared of grass.