Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP222
Serial No:K1350103
Tag No:
Acquisition Date:2014-04-03
Project:DOE
Location on 2021-08-11:Lab,
Notes: 0-1% range

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
5412 2021-08-11 Tonzi-SoilCO2 CO2 Probe Tree2 Profile -0.04 Removed Lab none Removed from -4cm in the soil co2 profile near LiuKang's tree. This profile has not been running for a while due to a break in the power line. This style of measurements are being replaced by forced diffusion chambers.
3063 2017-09-14 Lab none Installed Tonzi CO2 Probe Tree2 Profile -0.04 Installed at Tonzi in the Tree 2 profile at -4cm after lab calibration. This is a new installation with new holes dug.
3022 2017-08-25 Vaira CO2 probe West Profile -0.02 Removed Lab none Removed from Vaira from West profile -2 cm position for calibration
2177 2016-02-23 Lab none Installed Vaira CO2 probe West Profile -0.02 Installed in the Vaira West profile at -2cm after lab calibration.
2151 2016-02-04 Vaira CO2 probe West Profile -0.02 Removed Lab none Removed from Vaira West profile at -2cm for calibration
2061 2014-12-09 Vaira CO2 probe West Profile -0.02 Location Vaira CO2 probe West Profile -0.02 Vaira West profile at -2cm
1450 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*1.022 + 63.48; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.