Equipment and Calibration History
History for Vaisala GMP222 K1350103
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5412 | 2021-08-11 | Tonzi-SoilCO2 CO2 Probe Tree2 Profile -0.04 | Removed | Lab none | Removed from -4cm in the soil co2 profile near LiuKang's tree. This profile has not been running for a while due to a break in the power line. This style of measurements are being replaced by forced diffusion chambers. |
3063 | 2017-09-14 | Lab none | Installed | Tonzi CO2 Probe Tree2 Profile -0.04 | Installed at Tonzi in the Tree 2 profile at -4cm after lab calibration. This is a new installation with new holes dug. |
3022 | 2017-08-25 | Vaira CO2 probe West Profile -0.02 | Removed | Lab none | Removed from Vaira from West profile -2 cm position for calibration |
2177 | 2016-02-23 | Lab none | Installed | Vaira CO2 probe West Profile -0.02 | Installed in the Vaira West profile at -2cm after lab calibration. |
2151 | 2016-02-04 | Vaira CO2 probe West Profile -0.02 | Removed | Lab none | Removed from Vaira West profile at -2cm for calibration |
2061 | 2014-12-09 | Vaira CO2 probe West Profile -0.02 | Location | Vaira CO2 probe West Profile -0.02 | Vaira West profile at -2cm |
1450 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.022 + 63.48; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |