Equipment and Calibration
View History
Equipment Entry
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
2707 | 2017-04-04 | Twitchell Rice DO Sensor | Removed | Lab none | Removed with Twitchell Rice tower decommissioning. |
2394 | 2016-08-12 | Lab none | Installed | Twitchell Rice DO Sensor | Reinstalled at the Rice tower after being removed for spring draw down and calibration in the lab.
Reconnected the sensor to port 9, reading 27mV when connected and immersed in water in well. DO sensor was ziptied about 6-8 inches from the top of the well. |
2390 | 2016-08-11 | Lab none | Serviced | Lab none | Scrubbed with vinegar to remove residue and replaced small o-ring and membrane.
The new calibration is: M = 8.73/17.8 in (mg/L)/mV (aka ppm/mV) (@ 22C). |
2238 | 2016-04-13 | Twitchell Rice DO Sensor | Removed | Lab none | Removed with tower from field for planting. Needs to be serviced in the lab and reinstalled when the field is flooded. |
2114 | 2015-11-25 | Lab none | Installed | Twitchell Rice DO Sensor | Reinstalled at the Rice tower after being removed for fall draw down. |
2102 | 2015-11-23 | Lab none | Serviced | Lab none | Serviced probe to go back in the field this week. The new calibration is:
M = 9.45/36.1 in (mg/L)/mV (aka ppm/mV) (@ 18C). |
1857 | 2015-05-19 | Lab none | Calibration | Lab none | Serviced probe to go back in the field this week. The new calibration is:
M = 9.45/39.8 in (mg/L)/mV (aka ppm/mV) (@ 18C). |
1743 | 2015-03-05 | Twitchell Rice DO Sensor | Removed | Lab none | Removed from Twitchell Rice because the field is dry and we moved the tower for plowing |
1628 | 2014-11-11 | Lab none | Installed | Twitchell Rice DO Sensor | Re-installed at the rice. |
1627 | 2014-11-06 | Lab none | Serviced | Lab none | Serviced probe to back back in the field next week. The new calibration is:
M = 8.24/28.61 in (mg/L)/mV (aka ppm/mV) (@ 25C). |
1558 | 2014-09-10 | Twitchell Rice DO Sensor | Removed | Lab none | Removed from the Twitchell rice because the field in drained. |
1346 | 2014-06-12 | Lab none | Installed | Twitchell Rice DO Sensor | Installed at the rice. DO-08 was calibrated in the field: M = 8.24/38.61 in (mg/L)/mV (aka ppm/mV) (@ 23C). The probe is 3 cm above the soil surface. |