Rec No. | Date | Loc From | Title | Loc To | Notes |
2225 |
2016-04-05 |
Lab none |
Installed |
Vaira |
Reinstalled at Vaira in the Peltier SHF plate profile at -2cm after fixing broken wire in the lab. |
2222 |
2016-03-23 |
Vaira |
Removed |
Lab none |
Removed from Vaira at -2cm due to broken wire. |
2129 |
2015-12-21 |
Lab none |
Repair |
Lab none |
The red lead looked chewed and was broken at its junction to the plate. I soldered a new red wire directly to the junction and a new black wire about 3cm from the plate. Both wires were covered with heat shrink and the junctions to the plate were painted with black Rustoleum paint. |
2105 |
2015-12-02 |
Vaira |
Removed |
Lab none |
Removed from Vaira at -2cm. The red wire was broken not clear whether corrosion or animal (or shovel) damage |
915 |
2013-02-06 |
Lab none |
Installed |
Vaira |
Installed in a profile at -2cm to the east of the Vaira radiation tower |
920 |
2013-01-16 |
Lab none |
Calibration |
Lab none |
Calibrated by Sebastian at University of British Columbia: Multiplier = 99.54 (W/m2)/mV |