Equipment and Calibration History
History for Gill WM 1590 120908
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
7002 | 2024-10-24 | Lab none | Installed | Rush Ranch Anemometer | Installed on new mount at the top of the main tower at Rush Ranch |
6590 | 2024-02-28 | West Pond Anemometer | Removed | Lab none | Removed from WP as part of its decomissioning |
2581 | 2017-01-17 | Lab none | Installed | West Pond Anemometer | Installed on West Pond tower to replace sn 120907. |
2572 | 2017-01-06 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug.
Old SN: Y120908 New SN: X120908 Upgrade file: Y074417-Y124215-m.gcu, Y012345-s.gcu, X120908 X124215-m.gcu Old firmware: 2329-114 New firmware: 2329-116-01 All Checksum fields report PASS. There are multiple upgrade .gcu files. During the initial upgrade using file "Y074417-Y124215-m.gcu," the sensor was set at 57,600 baud rate. This baud rate caused an error in the upgrade, so the SN reset to Y012345. Gill sent me the second upgrade file, "Y-12345-s.gcu," which reset the sensor to SN X120908. Finally, I used the third upgrade file, "X120908 X124215-m.gcu," to upgrade the firmware. Log file: WMUpgradeLog_X120908_170106_122011.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Baud Rate : 19200 Baud Rate : 57600 Output rate : P1 1 Hz Output rate : P8 20 Hz THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X120908 ---- SW VERSION-------------------------------------------- Anemometer Firmware Version: 2329-116 ---- HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Baud Rate : 57600 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : SOS Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound |
2147 | 2016-01-12 | Twitchell Alfalfa Anemometer | Removed | Lab none | Removed from field for calibration |
1006 | 2013-05-24 | Lab none | Installed | Twitchell Alfalfa Anemometer | Installed on the northwest corner of the scaffolding at the Alfalfa site. |
855 | 2012-10-30 | Mayberry-Portable | Removed | Lab none | Removed from the portable tower and returned to the lab. |
854 | 2012-05-10 | Lab none | Installed | Mayberry-Portable | Initial installation on the Protable Tower at Mayberry. |