History for Campbell Sci. NL115 9774

Rec No.DateLoc FromTitleLoc ToNotes
5048 2016-08-12 Lab none Installed Bouldin Alfalfa none Installed at BA with initial site set up
1102 2013-11-13 Factory Returned Lab none Returned from Campbell. They replaced the bad eject button and updated the firmware. No charge - under warranty.
1087 2013-11-04 Lab none Shipped Factory Sent to Campbell for a check on the non-functioning eject button on RMA 18887.
1086 2013-10-18 Tonzi-Floor Removed Lab none Removed from the Floor Tower. The eject button was not giving a green light to eject the card and the files were getting corrupted on a forced eject. I brought the module in for testing. Replace by sn 6761
809 2012-10-09 Lab none Installed Tonzi-Floor Installed with CR1000 sn 50889 to collect flux data at the Floor tower in parallel with winflux. This interface is in the box with the sonic PSU.