Equipment and Calibration

View History

Equipment Entry

Instrument:PAR Sensor
Manufacture:Kipp & Zonen
Model No:PAR-LITE
Serial No:030381
Tag No:
Acquisition Date:2004-02-10
Project:DOE-TCP
Location on 2020-09-16:Vaira,
Notes: Field calibration 5.43 uV/(umol/m2/s) 2019-06-18

Documents

Calibration: cal_KippZonen_PARLite_sn030381_20031217.pdf
Report: PAR comparison 12-2018 - 2018-12-11
Report: Oxford Tract Radiometer Comparison - 06/2019 - 2019-06-18
Report: Vaira PAR Cross Compare - 2020-06-12

History

Rec No.DateLoc FromTitleLoc ToNotes
4800 2020-09-16 Vaira PAR Sensor Reflected Cal Moved Vaira PAR Sensor Reflected Moved from the reflected cal position to the main reflected position.
4711 2020-08-26 Lab none Installed Vaira PAR Sensor Reflected Cal Installed at Vaira in the reflected position for inter-comparison. This sensor will likely replace the old sensor.
4712 2020-08-05 Bouldin Corn PAR Sensor Under-B Removed Lab none Removed from under the canopy at Bouldin Corn. The corn has reached maximum growth continued measurements of questionable use.
4510 2020-05-29 Lab none Installed Bouldin Corn PAR Sensor Under-B I stalled about 10cm above the ground near the picams to measure under canopy PAR
4071 2019-06-18 Lab none Calibration Lab none New calibration factor of 5.43 uV/(umol/m2/s). Tested this sensor at Oxford Tract against AMP reference sensor K&Z PQS1 sn 161073. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details.
3775 2018-10-30 Lab none Location Lab none This sensor has been on the tram in the outgoing position for at least the last 10 years.
4070 2003-12-17 Factory Calibration Lab none Initial factory calibration 5.32 uV/(umol/m2/s)