Equipment and Calibration
View History
Equipment Entry
|
Documents
Calibration: cal_KippZonen_PARLite_sn030381_20031217.pdfReport: PAR comparison 12-2018 - 2018-12-11
Report: Oxford Tract Radiometer Comparison - 06/2019 - 2019-06-18
Report: Vaira PAR Cross Compare - 2020-06-12
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
4800 | 2020-09-16 | Vaira PAR Sensor Reflected Cal | Moved | Vaira PAR Sensor Reflected | Moved from the reflected cal position to the main reflected position. |
4711 | 2020-08-26 | Lab none | Installed | Vaira PAR Sensor Reflected Cal | Installed at Vaira in the reflected position for inter-comparison. This sensor will likely replace the old sensor. |
4712 | 2020-08-05 | Bouldin Corn PAR Sensor Under-B | Removed | Lab none | Removed from under the canopy at Bouldin Corn. The corn has reached maximum growth continued measurements of questionable use. |
4510 | 2020-05-29 | Lab none | Installed | Bouldin Corn PAR Sensor Under-B | I stalled about 10cm above the ground near the picams to measure under canopy PAR |
4071 | 2019-06-18 | Lab none | Calibration | Lab none | New calibration factor of 5.43 uV/(umol/m2/s). Tested this sensor at Oxford Tract against AMP reference sensor K&Z PQS1 sn 161073. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
3775 | 2018-10-30 | Lab none | Location | Lab none | This sensor has been on the tram in the outgoing position for at least the last 10 years. |
4070 | 2003-12-17 | Factory | Calibration | Lab none | Initial factory calibration 5.32 uV/(umol/m2/s) |