Equipment and Calibration History
History for Gill WM 1590 074417
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5380 | 2021-07-20 | Lab none | Installed | Bouldin Alfalfa Anemometer | Installed at Bouldin Alfalfa to replace sn 152806 for advection experiemnt. This sonic was compare to the other three used for the advection experiement. |
4411 | 2020-03-23 | Factory none | Returned | Lab none | Returned from Novanlynx and Gill - rebuilt from spare parts due to lost transducer, cracked spar and internal water damage. Maybe more detail on packing slip? |
4329 | 2020-01-21 | Lab | Shipped | Factory | Shipped to Novalynx which will send it on to Gill for repair of the missing transducer and test to Specification. |
4276 | 2019-12-02 | Bouldin Corn Anemometer | Removed | Lab none | One traducer missing. Removed from Bouldin Corn site and brought back to the lab after the tower fell down during a rain/wind storm on 2019-11-26. |
2717 | 2017-04-26 | Lab none | Installed | Bouldin Corn Anemometer | Installed at 5.2m agl at the Bouldin Corn site with the initial set up of the tower. |
2621 | 2017-02-17 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug. The upgrade failed the first time because the Power-on message had been turned off. Without the Power-on message, the upgrade wizard cannot check if the checksums are correct. After I turned on the Power-on message, I used the second upgrade (X074417-s.gcu) file to complete the upgrade. Old SN: Y074417 New SN: X074417 Upgrade file: Y074417-Y124215-m.gcu, X074417-s.gcu Old firmware: 2329-114-01 New firmware: 2329-116-01 All Checksum fields report PASS. Log file: WMUpgradeLog_X074417_170217_121544.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Output rate : P2 2 Hz Output rate : P8 20 Hz THE NEW CONFIGURATION IN FULL ====DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X074417 ----SW VERSION-------------------------------------------- Anemometer Firmware Version: 2329-116 ----HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 ADC resolution of attached unit : 14 DAC resolution of attached unit : 14 Number of ADC channels : 4 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ----USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P2 2 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Fixed Field SOS and Sonic Temperature display : SOS Analogue input display : OFF Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound |
2596 | 2017-01-26 | Mayberry Anemometer | Removed | Lab none | Removed from Mayberry tower so we could take it back to the lab for firmware upgrade. It was replaced by sn 153804. |
548 | 2012-04-24 | Lab none | Installed | Mayberry Anemometer | Installed on the Mayberry eddy tower the new Licor GHG system to swap out WMP sonic sn 000019. The old sonic did not have analog output, the new one does. |
1675 | 2012-02-25 | Factory | Returned | Lab none | Returned from factory upgrade through Dynamax. 4x14 bit analog outputs were added. |
1674 | 2012-01-05 | Lab none | Shipped | Factory | Shipped to Dynamax to add 4x14 bit analog outputs. |
352 | 2009-09-17 | Factory | Returned | Lab none | Returned from PP Systems for and upgrade of the firmware to fix smoothing problem with air temperature. |
1962 | 2009-06-26 | Sherman-2ndTower | Removed | Lab none | Removed from the secondary tower at Sherman. This sonic has a bug that smoothes the SOS or T channel. It will be sent to Gill for a firmware upgrade. |
1961 | 2008-11-25 | Lab none | Installed | Sherman-2ndTower | Installed on a secondary tower about 1m north of the main tower. Installed with a 7500 and CR1000. |