Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP221
Serial No:Y1720006
Tag No:
Acquisition Date:0000-00-00
Project:Tonzi
Location on 2014-05-27:Lab,
Notes: 0-2% range

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
1461 2014-05-27 Lab none Calibration Lab none Trueppm = Probeppm*0.6676 + -1028; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine.
2008 2014-04-22 Vaira 0 Removed Lab none Removed from Vaira West profile at -4cm, replaced by sn K1310102
2020 2014-03-18 Vaira CO2 probe West Profile -0.08 Moved Vaira CO2 probe West Profile -0.02 At Vaira in the West profile this probe was moved from the -8cm location to -2cm
2004 2014-01-29 Lab none Installed Vaira CO2 probe West Profile -0.08 West profile as -8cm with new U-shaped tops on the PVC enclosures.
1441 2014-01-27 Lab none Calibration Lab none Trueppm = Probeppm*0.9255 + -857.4; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty.
1195 2014-01-24 Vaira CO2 probe West Profile -0.08 Removed Lab none Removed from Vaira west plot at -8cm for lab calibration
1129 2012-03-21 Lab none Installed Vaira CO2 probe West Profile -0.08 Installed at Vaira in the west set at -8cm
1413 2012-02-06 Lab none Calibration Lab none Trueppm = Probeppm*1.08 + 78.35; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan.
1404 2012-02-03 Lab none Calibration Lab none Trueppm = Probeppm*0.9432 + -643; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.6C, P=101.6kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan.
1292 2010-04-06 Lab none Installed Sherman Island Installed at Sherman Island in Set A at -32cm.
1357 2010-02-03 Lab none Calibration Lab none Trueppm = Probeppm*1.011 + -103.2; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, 597ppm, and 2238ppm. Ta=24.1C, P=101.2kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Jaclyn.
12 2009-12-01 Factory Repair Factory Vaisala factory repair and calibration 0-2% - certificate on file.
Problem: Got probe error 05, 10, and 11. No IR source and memory error.
Action: Replaced probe circuit board and adjusted unit.