Field Notes

<--2024-06-27 13:10:00 | 2024-07-22 12:00:00-->
Other sites visited today: Bouldin Corn | East End | Gilbert Tract
Phenocam link: Searching...

Bouldin Alfalfa: 2024-07-17 09:45:00 (DOY 199)
Author: Daphne Szutu
Others: Robert

Summary: Updated CR1000 datalogger program to edit modified Bowen calc, FD chamber performance check, removed PRIout sensor

2024-07-17
Bouldin Alfalfa

Robert and I arrived at 9:45 PST (10:45 PDT). It was sunny and mild with a cool breeze. The first field on the left when we turn off of the highway has been planted with corn this year (homogenous corn canopy). Our alfalfa field was flowering with lots of white butterflies flitting about.

Robert took veg height and spectral measurements and cleaned the rad and flux sensors.

I downloaded met, cam, USB GHG, and FD data.

I moved the FD chamber to lay on its side and ran an annual routine performance check on the FD chamber. I accidentally lost connection to the chamber about halfway through (jiggled the USB-ethernet dongle when I shifted position), so I had to start over. Once it finished, I uploaded the FD offset and exported the results. I installed the FD chamber back in its usual vertical orientation.

I updated the CR1000 datalogger program to edit the modified Bowen-variance method calculations. It now uses a constant pressure instead of no air pressure data to calculate qsat. I also removed the code for the PRI sensors (narrow FOV PRI sensor was still reading, but incoming PRI sensor had been removed for a while). I removed the PRIout sensor sn 930103961 from the rad boom and brought it back to the lab.

While I was waiting for the FD to finish its performance check, I tightened the datalogger and AM16/32 screws. All seemed fine. I also washed the solar panels and trimmed some of the taller weeds around the panels.

The 7500 read: 404ppm CO2, 666mmol/m3 H2O, 22.7C, 101.6kPa, 98SS—99 after cleaning

We left at 10:50 PST (11:50 PDT). Patty and I talked on the phone a few more times about her program. They confirmed that the wiring was correct and that both the 7200 and the CSAT3B were working and producing reasonable numbers, but they weren’t seeing anything on the datalogger for the microSD card. She checked that the serial addresses of the sensor matched the datalogger program. In the end, she said they were going back inside for a break from the heat and would call Stephen for help.

Back in the lab, I took the spectral card out at some time and couldn't find it again. No spectral data from this date for now.

Next Mowing: 2024-07-26

Veg Height (cm): Tape measure from ground to top of plant
44, 48, 43, 32, 42, 53, 39, 28, 30, 26, avg = 38.5, stdev = 9.1

0 photos found

Graphs display:
7 sets found

4 BA_flux graphs found

Explore the graph:Wind in a new window

Explore the graph:LI-7500 CO2 and H2O in a new window

Explore the graph:Counts in a new window

Explore the graph:CO2 in ppm in a new window

21 BA_met graphs found

Explore the graph:Soil Temperatures Set A in a new window

Explore the graph:Soil Temperatures Set B in a new window

Explore the graph:Soil Temperatures Set C in a new window

Explore the graph:Soil Heat Flux in a new window

Explore the graph:Temperature in a new window

Explore the graph:Solar Radiation in a new window

Explore the graph:Air Pressure and RH in a new window

Explore the graph:Soil Moisture in a new window

Explore the graph:PAR in a new window

Explore the graph:PRI Channels in a new window

Explore the graph:NDVI channels in a new window

Explore the graph:Battery Voltage in a new window

Explore the graph:NDVI and PRI in a new window

Explore the graph:Apogee NDVI in a new window

Explore the graph:FD Chamber A CO2 and Temp in a new window

Explore the graph:FD Chamber B CO2 and Temperature in a new window

Explore the graph:Surface Temperature in a new window

Explore the graph:Sentek Soil Moisture profile in a new window

Explore the graph:Sentek Soil Temperature profile in a new window

Explore the graph:Sentek Soil Salinity profile in a new window

Explore the graph:Forced Diffusion Soil CO2 Flux in a new window

6 BA_arable graphs found

Explore the graph:Incoming Solar Bands in a new window

Explore the graph:Reflected Solar Bands in a new window

Explore the graph:Solar Radiation in a new window

Explore the graph:Temperatures, PPT in a new window

Explore the graph:Sentek Deep Soil Moisture in a new window

Explore the graph:Pressure, RH in a new window

Can't check TATWBABCSB_cam
Can't check MBSWEEWPBCBA
2 BA_processed graphs found

Explore the graph:Energy in a new window

Explore the graph:Evapotranspiration in a new window

No data found for BA_adex.