Field Notes
<--2022-08-16 13:10:00 | 2022-09-15 13:00:00-->Other sites visited today: Gilbert Tract | East End | Mayberry
Phenocam link: Searching...
West Pond: 2022-08-30 13:10:00 (DOY 242)
Author: Daphne Szutu
Others: Ari
Summary: Regular data collection/cleaning, water level going down, spiderwebs on 7500--cleaned
2022-08-30 West Pond Ari and I arrived at 13:10 PDT. It was sunny, warming up, and muggy among the plants. Water level is decreasing. The staff gauge read 41cm. The conductivity sensor was out of the water, so I pulled more slack out of the extra coiled cable and dropped the sensor into the water under the lower boardwalk by the staff gauge. Fewer reeds there to impede the sensor. I downloaded met, cam, and USB GHG data. There were small ants and a pile of their eggs in the met box so I left them some mothballs. Ari cleaned flux and rad. There were a bunch of spiderwebs in the 7500, so the SS was a low 73. CO2/H2O numbers looked better after cleaning. We topped off the wash reservoir. There were quite a few wasps in the reservoir box so I just left the lid open. The 7500 read: 360ppm CO2, 1000mmol/m3 H2O, 28.7, 101.3kPa, 73SS After cleaning, the 7500 read: 400ppm CO2, 778mmol/m3 H2O, 29,1C, 101.3kPa, 102SS The 7700 read: 2.0ppm CH4, 71RSSI – 83 after cleaning We left at 13:35 PDT. |
0 photos found
8 sets found
Can't check WP_DO
1 WP_DO graphs found
14 WP_met graphs found
5 WP_flux graphs found
Can't check MBWPEESWGT_mixed
Can't check MBEEWPSWEPGT_cam
Can't check MBSWEEWPBCBA
1 WP_processed graphs found
4 WP_arable graphs found