Field Notes
<--2018-01-16 11:50:00 | 2018-01-31 10:15:00-->Other sites visited today: Mayberry | East End | Twitchell Alfalfa
Phenocam link: Searching...
West Pond: 2018-01-23 13:15:00 (DOY 23)
Author: Joe Verfaillie
Others: Daphne
Summary: Muddy roads, checked on GHG clocks.
West Pond 2018-01-23 Daphne and I arrived at 13:15. It was cool and mostly sunny with clouds on the horizon. The road between the ponds and north of the ponds is soft and muddy. This system has been missing a few 7700 reading approximately hourly. It was showing a consistent "time sync" yellow triangle error. We set the 7700 clock to automatic and reset the 7550 box. The "time sync" error disappeared. Daphne collected the GHG data and did not see any missing data. She also double checked the columns being collected from the 7700 – seemed okay. The met data looked okay too. The 7550 box here needs some TLC. The gasket is a mess of holes and putty, the pressure port is missing (hole is covered with electrical tape), the reset button is missing and the status LEDs don’t seem to be aligned with the holes in the cover. We left at 13:30 |
0 photos found
8 sets found
Can't check WP_DO
1 WP_DO graphs found
14 WP_met graphs found
5 WP_flux graphs found
Can't check MBWPEESWGT_mixed
Can't check MBEEWPSWEPGT_cam
Can't check MBSWEEWPBCBA
1 WP_processed graphs found
No data found for WP_arable.