Instrument: | pH ORP |
Manufacture: | YSI |
Model No: | EXO pH-ORP |
Serial No: | 24G102326 |
Tag No: | |
Acquisition Date: | 2024-07-25 |
Project: | DeltaCouncil-Dutch |
Location on 2024-10-23: | Gilbert Tract, |
Notes: |
Electrolyte in the sensor module will be depleted in 12-18 months at which point the module will need to be replaced.
Ideally, delta slope > 165 mV, and the mV per decade = 59.16 +/- 5.
Part number 577612
0 to 14 unit measurement range
T63<3 sec response time for pH, T63<5 response time for ORP
±0.1 pH unit accuracy within ±10°C of calibration temperature
±20 mV ORP unit accuracy in Redox standard solutions
|
|
|